NAMIC Convenience Kit, RADIOLOGY KIT, UPN H7496002072111, REF/Catalog No. 600207211, STERILE, Rx ONLY --- For single use only. Device Listing Number D136683. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.

Published March 4, 2015 · Recall initiated: January 9, 2015

Summary

FDA records medical device Class II recall Z-1207-2015, published March 4, 2015 and initiated January 9, 2015. Recalling firm: Navilyst Medical, Inc; Status: Terminated; Reported quantity: 133 units; Lot: Lots 4244975, 4251615, 4269276, 4292593, 4310081, 4344776, 4374583, 4384503, 4413658, 4447492, 4505410, 4535348, 4569287, 4572788, 4604137, 4610092, 4612832, 4630601, 4630602, 4638044, 4681013, 4690175, 4705330, 4717942, 4731420, 4772544, 4812433. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Navilyst Medical, Inc
Distributor
Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
Lot
Lots 4244975, 4251615, 4269276, 4292593, 4310081, 4344776, 4374583, 4384503, 4413658, 4447492, 4505410, 4535348, 4569287, 4572788, 4604137, 4610092, 4612832, 4630601, 4630602, 4638044, 4681013, 4690175, 4705330, 4717942, 4731420, 4772544, 4812433.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was NAMIC Convenience Kit, RADIOLOGY KIT, UPN H7496002072111, REF/Catalog No. 600207211, STERILE, Rx ONLY --- For single use only. Device Listing Number D136683. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems. recalled?

Yes. FDA published recall Z-1207-2015 on March 4, 2015. Reported hazard: The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

What should I do if I have this product?

Voluntary: Firm initiated · Letter