Gyrus ACMI Disposable Falope-Ring Band Applicator Kits. REF 006889-901 and REF 005280-901 (with 8 mm Disposable Trocan). Sterile EO. Single Use Only. The disposable Falope-Ring Band Applicator is a sterile, two-band, dual-incision instrument. The two-band applicator enables the physician to occlude both Fallopian tubes without removing the applicator from the abdominal cavity. The Falope-Ring Band and Applicator Kits are indicated for use for female sterilization (permanent contraception).
Published March 19, 2014 · Recall initiated: February 19, 2014
Summary
FDA records medical device Class II recall Z-1196-2014, published March 19, 2014 and initiated February 19, 2014. Recalling firm: Gyrus Medical, Inc; Status: Terminated; Reported quantity: 856 units; Lot: Lots: MK761655, MK764022, MK767655, MK526401. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to an anomaly in the packaging process for specific lots of Disposable Falope-Ring Band Kit, it is possible that a defective seal could be present that may allow a breach of the package's sterile barrier and may compromise the sterility of the product. The breach may or may not be easily seen.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Gyrus Medical, Inc
- Distributor
- US Distribution.
- Lot
- Lots: MK761655, MK764022, MK767655, MK526401
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGYRUS ACMI, REF 3345 -Everest BiCOAG MOLly Forceps 5mm/45cm REF 3345PK - PKS MOLly Forceps 5mm/45cm REF 3640October 29, 2014
- FDAGyrus ACMI Falope-Ring Dilator, item 000878-4. The Falope-Ring Dilator is a conical plastic device that is aAugust 27, 2014
- FDACELLTRACKS ANALYZER II® (Product Code 9555) Product Usage: The CELLTRACKS ANALYZER II® is a semi-automatedMarch 19, 2014
- FDAXiO RPT System The XiO RTP System is used to create treatment plans for any cancer patient for whom externalMarch 19, 2014
- FDAPhilips HeartStart MRx Monitor/Defibrillator Models: M3535A and M3536A with Q-CPR Meter Option B08 TheMarch 19, 2014
- FDAGYRUS ACMI, REF 3740, Everest BiCOAG Bipolar Dissecting Forceps 5mm/45 cm, Sterile EO. GYRUS ACMI, REF 3740PK,August 28, 2013
Questions about this recall
Was Gyrus ACMI Disposable Falope-Ring Band Applicator Kits. REF 006889-901 and REF 005280-901 (with 8 mm Disposable Trocan). Sterile EO. Single Use Only. The disposable Falope-Ring Band Applicator is a sterile, two-band, dual-incision instrument. The two-band applicator enables the physician to occlude both Fallopian tubes without removing the applicator from the abdominal cavity. The Falope-Ring Band and Applicator Kits are indicated for use for female sterilization (permanent contraception). recalled?
Yes. FDA published recall Z-1196-2014 on March 19, 2014. Reported hazard: Due to an anomaly in the packaging process for specific lots of Disposable Falope-Ring Band Kit, it is possible that a defective seal could be present that may allow a breach of the package's sterile barrier and may compromise the sterility of the product. The breach may or may not be easily seen.
What should I do if I have this product?
Voluntary: Firm initiated · Letter