Colonoscope, Model Number PCF-H190L.
Published March 6, 2024 · Recall initiated: January 11, 2024
Summary
FDA records medical device Class II recall Z-1211-2024, published March 6, 2024 and initiated January 11, 2024. Recalling firm: Olympus Corporation of the Americas; Status: Ongoing; Reported quantity: 31; Lot: Model Number: PCF-H190L; UDI DI: 04953170305191; Serial Number: 2201678, 2201735, 2301759, 2301775, 2301929, 2301930, 2302159, 2302280, 2302281, 2302380, 2302506, 2303832, 2303927, 2404124, 2404132, 2404806, 2404888, 2404967, 2405012, 2405253, 2405290, 2405353, 2517013, 2517120, 2517422, 2517440, 2517541, 2517616, 2517649, 2517705, 2628544. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Olympus Corporation of the Americas
- Distributor
- US Nationwide distribution.
- Lot
- Model Number: PCF-H190L; UDI DI: 04953170305191; Serial Number: 2201678, 2201735, 2301759, 2301775, 2301929, 2301930, 2302159, 2302280, 2302281, 2302380, 2302506, 2303832, 2303927, 2404124, 2404132, 2404806, 2404888, 2404967, 2405012, 2405253, 2405290, 2405353, 2517013, 2517120, 2517422, 2517440, 2517541, 2517616, 2517649, 2517705, 2628544;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOlympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.October 7, 2026
- FDAOlympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.October 7, 2026
- FDAOlympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunctionOctober 7, 2026
- FDAOlympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.September 30, 2026
- FDAColonoscope, Model Number PCF-H190DL.March 6, 2024
- FDAColonoscope, Model Number PCF-H190TL.March 6, 2024
Questions about this recall
Was Colonoscope, Model Number PCF-H190L. recalled?
Yes. FDA published recall Z-1211-2024 on March 6, 2024. Reported hazard: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
What should I do if I have this product?
Voluntary: Firm initiated · Letter