Natural-Knee® II Constrained Knee System NKII CCK TIB INS SZ00/0-L NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ3-5 -L NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ3-5 -R NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ3-5 -L NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ1/2-RT NKII CCK TIB INS SZ3-5 -R NKII CCK TIB INS SZ00/0-L NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ3-5 -L NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ3-5 -R NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ3-5 -L NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ3-5 -R NKII CCK TIB INS SZ00/0-L NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ1/2-RT NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ3-5 -L NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ1/2-RT NKII CCK TIB INS SZ00/0-L NKII CCK TIB INS SZ00/0-R
Published March 26, 2014 · Recall initiated: February 25, 2014
Summary
FDA records medical device Class II recall Z-1256-2014, published March 26, 2014 and initiated February 25, 2014. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 3638 units; Lot: Item Number 672009100 672009101 672009103 672009200 672009203 672011101 672011103 672011200 672011201 672011203 672013100 672013101 672013103 672013200 672013203 672016101 672016103 672016200 672016203 672019100 672019101 672019200 672019201 672022101 672022103 672022200 672022201 672025100 672025200. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
- Lot
- Item Number 672009100 672009101 672009103 672009200 672009203 672011101 672011103 672011200 672011201 672011203 672013100 672013101 672013103 672013200 672013203 672016101 672016103 672016200 672016203 672019100 672019101 672019200 672019201 672022101 672022103 672022200 672022201 672025100 672025200
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDANatural-Knee® II System ULTRA CONG INS LT SZA/9MM ULTRA CONG INS LT SZA/13M ULTRA CONG INS LT SZA/19M ULTRAMarch 26, 2014
- FDANatural-Knee® II Posterior Stabilized (P.S.) Condylar Tibial Inserts: N-K II CONS PS TIB INS SZ N-K II CONS PSMarch 26, 2014
Questions about this recall
Was Natural-Knee® II Constrained Knee System NKII CCK TIB INS SZ00/0-L NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ3-5 -L NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ3-5 -R NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ3-5 -L NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ1/2-RT NKII CCK TIB INS SZ3-5 -R NKII CCK TIB INS SZ00/0-L NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ3-5 -L NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ3-5 -R NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ3-5 -L NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ3-5 -R NKII CCK TIB INS SZ00/0-L NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ1/2-RT NKII CCK TIB INS SZ1/2-LF NKII CCK TIB INS SZ3-5 -L NKII CCK TIB INS SZ00/0-R NKII CCK TIB INS SZ1/2-RT NKII CCK TIB INS SZ00/0-L NKII CCK TIB INS SZ00/0-R recalled?
Yes. FDA published recall Z-1256-2014 on March 26, 2014. Reported hazard: A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
What should I do if I have this product?
Voluntary: Firm initiated