Back to search
FDAFDA · Food and drugs·Recall number Z-1260-2014

Apollo" Knee System: PS TIB INS SZ0/16MM APOLL PS TIB INS SZ0/19MM APOLL PS TIB INS SZ0/22MM APOLL PS TIB INS SZ1/16MM APOLL PS TIB INS SZ2/16M APOLLO PS TIB INS SZ2/22MM APOLL PS TIB INS SZ3/16MM APOLL PS TIB INS SZ3/19MM APOLL PS TIB INS SZ4/9MM APOLLO PS TIB INS SZ4/11MM APOLL PS TIB INS SZ4/13MM APOLL PS TIB INS SZ4/16MM APOLL PS TIB INS SZ4/19MM APOLL PS TIB INS SZ4/22MM APOLL PS TIB INS SZ5/9MM APOLLO PS TIB INS SZ5/19MM APOLL

Published March 26, 2014 · Recall initiated: February 25, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Lot
Item Number Item Number 681600016 681600019 681600022 681601016 681602016 681602022 681603016 681603019 681604009 681604011 681604013 681604016 681604019 681604022 681605009 681605019
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card