FDAFDA · Food and drugs·Recall number Z-1261-2014
Apollo" Revision/Constrained Knee System: SZ0-16MM AK CCK TIB INSER SZ1-11MM AK CCK TIB INSER SZ1-16MM AK CCK TIB INSER SZ1-22MM AK CCK TIB INSER SZ1-30MM AK CCK TIB INSER SZ2-11MM AK CCK TIB INSER SZ2-13MM AK CCK TIB INSER SZ2-16MM AK CCK TIB INSER SZ2-22MM AK CCK TIB INSER SZ2-30MM AK CCK TIB INSER SZ3-16MM AK CCK TIB INSER SZ3-22MM AK CCK TIB INSER SZ3-25MM AK CCK TIB INSER SZ4-13MM AK CCK TIB INSER SZ4-16MM AK CCK TIB INSER SZ5-11MM AK CCK TIB INSER
Published March 26, 2014 · Recall initiated: February 25, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
What to do
Voluntary: Firm initiated
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
- Lot
- Item Number 682000016 682001011 682001016 682001022 682001030 682002011 682002013 682002016 682002022 682002030 682003016 682003022 682003025 682004013 682004016 682005011
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.