The E.CAM gamma camera system Product Usage: The E.CAM gamma camera system is a SPECT system and measures the distribution of radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body
Published April 6, 2016 · Recall initiated: February 26, 2016
Summary
FDA records medical device Class II recall Z-1262-2016, published April 6, 2016 and initiated February 26, 2016. Recalling firm: Siemens Medical Solutions USA, Inc.; Status: Terminated; Reported quantity: 3090 units; Lot: 10275879, 10413009, 04380213, 04380221, 05242826, 05977066, 05977074, 05984005, 05989079, 05989087, 05989095, 05991109, 05991117, 05992099, 07324119, 07324135, 07324143, 07324150, 07760809, 07760932, 07761161, 07823920, 07823938, 07823946, 07823953, 07823961, 07823979, 10151531, 10151532, 10520745. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
We received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-00002). The result of a Hazard Analysis and CAPA investigation, we have determined that for e.cam or Symbia E systems that although the existing user manual is correct, additional supplemental instructions, with the inclusion of visual aids, may further clarify and emphasize the proper collimator change
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Distributor
- Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Puerto Rico.***** Foreign Consignees: Albania, Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, El Salvador, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kosovo, Kuwait, Libya, Malaysia, Mauritius, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Oman, P.R. China, Pakistan, Palestinian Ter, Peru, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad,Tobago, Tunisia, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay and Vietnam.
- Lot
- 10275879, 10413009, 04380213, 04380221, 05242826, 05977066, 05977074, 05984005, 05989079, 05989087, 05989095, 05991109, 05991117, 05992099, 07324119, 07324135, 07324143, 07324150, 07760809, 07760932, 07761161, 07823920, 07823938, 07823946, 07823953, 07823961, 07823979, 10151531, 10151532, 10520745
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with softwareOctober 2, 2024
- FDAACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON JuniperOctober 2, 2024
- FDAACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.0October 2, 2024
- FDAACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10GMay 1, 2024
- FDA30" (76 cm) Appx. 6.3 ml, 20 Drop Admin Set w/ Integrated ChemoLock Port Drip Chamber, 0.2 Micron Filter,April 6, 2016
- FDAThe Symbia gamma camera system Product Usage: The Symbia gamma camera system is a computed tomography x-rayApril 6, 2016
Questions about this recall
Was The E.CAM gamma camera system Product Usage: The E.CAM gamma camera system is a SPECT system and measures the distribution of radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body recalled?
Yes. FDA published recall Z-1262-2016 on April 6, 2016. Reported hazard: We received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-00002). The result of a Hazard Analysis and CAPA investigation, we have determined that for e.cam or Symbia E systems that although the existing user manual is correct, additional supplemental instructions, with the inclusion of visual aids, may further clarify and emphasize the proper collimator change
What should I do if I have this product?
Voluntary: Firm initiated · Letter