ACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON Juniper Select 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11653093, with software.
Published October 2, 2024 · Recall initiated: August 15, 2024
Summary
FDA records medical device Class II recall Z-3215-2024, published October 2, 2024 and initiated August 15, 2024. Recalling firm: Siemens Medical Solutions USA, Inc.; Status: Ongoing; Reported quantity: 7022; Lot: 11335791, UDI-DI: 04056869152059: 1.0/VA10G and prior, 1.5/ VB10H and prior , 2.0/ VB11C and prior, 2.5/ VB30D and prior. 11653093, UDI-DI: 04056869957357: 2.0/ VB11C and prior, 2.5/ VB30D and prior. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Distributor
- US: MS, WA, TX, NJ, GA, PN, NY, AL, MN, NE, IN, CT, CA, IL, TN, WI, ND, NM, AR, NC, PR, RI, WY, MD, FL, HI, AZ, SC, MO, OK, VA, ME, KS, MI, LA, IA, OH, SD, NH, KY, AL, OR, MA, GU. OUS: Afghanistan, Albania, Algeria, Andorra, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Cameroon, Canada, Chile, China, Colombia, Congo, the Democratic Republic of the, Costa Rica, C®te dIvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guinea, French Polynesia, Georgia, Germany, Greece, Guatemala, Guernsey, Guyana, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein LI B, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Malta, Mexico, Moldova, Rep. of, Mongolia, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Reunion, Romania, Ruanda, Russian Federation, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe
- Lot
- 11335791, UDI-DI: 04056869152059: 1.0/VA10G and prior, 1.5/ VB10H and prior , 2.0/ VB11C and prior, 2.5/ VB30D and prior. 11653093, UDI-DI: 04056869957357: 2.0/ VB11C and prior, 2.5/ VB30D and prior.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with softwareOctober 2, 2024
- FDAACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.0October 2, 2024
- FDAGE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.October 2, 2024
- FDAprodisc C SK, Total cervical disc replacement, Model/Catalog Number: PDSM5October 2, 2024
- FDAACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10GMay 1, 2024
- FDAACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running softwareMay 1, 2024
Questions about this recall
Was ACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON Juniper Select 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11653093, with software. recalled?
Yes. FDA published recall Z-3215-2024 on October 2, 2024. Reported hazard: If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
What should I do if I have this product?
Voluntary: Firm initiated · Letter