i-STAT CG4+ cartridge (blue), List No. 03P85-50. UDI (01) 00054749002269 - Product Usage: The i-STAT CG4+ cartridge is intended for the quantitative measurement of lactate, pH, pCO2, and pO2 in venous, arterial or capillary whole blood.
Published February 26, 2020 · Recall initiated: January 15, 2020
Summary
FDA records medical device Class II recall Z-1263-2020, published February 26, 2020 and initiated January 15, 2020. Recalling firm: Abbott Point Of Care Inc.; Status: Terminated; Reported quantity: 99,850; Lot: All lots are affected. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and non-waived. Different CLIA standards apply to waived and non-waived tests as well as the facilities that perform these types of tests. The i-STAT CHEM8+ cartridge contains nine measured assays (sodium, potassium, chloride, blood urea nitrogen (BUN), ionized calcium, TCO2, glucose, creatinine and hematocrit) and is categorized as a CLIA waived test for venous whole blood samples, meaning this cartridge could be used in a facility with a CLIA Certificate of Waiver as a waived test.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Point Of Care Inc.
- Distributor
- US Nationwide distribution.
- Lot
- All lots are affected
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAi-STAT EG6+ cartridge; List Number: 03P77-25;May 6, 2026
- FDAi-STAT EG7+ cartridge; List Number: 03P76-25;May 6, 2026
- FDAi-STAT G3+ cartridge; List Number: 03P78-26;May 6, 2026
- FDAi-STAT EG6+ cartridge. List Number: 03P77-25.September 24, 2025
- FDAAtellica CH High Sensitivity C-Reactive Protein (HsCRP) Calibrator - Product Usage: For in vitro diagnosticFebruary 26, 2020
- FDAPilling Aortic Punch, 4.0 mm Standard Tip, REF 353440February 26, 2020
Questions about this recall
Was i-STAT CG4+ cartridge (blue), List No. 03P85-50. UDI (01) 00054749002269 - Product Usage: The i-STAT CG4+ cartridge is intended for the quantitative measurement of lactate, pH, pCO2, and pO2 in venous, arterial or capillary whole blood. recalled?
Yes. FDA published recall Z-1263-2020 on February 26, 2020. Reported hazard: The I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and non-waived. Different CLIA standards apply to waived and non-waived tests as well as the facilities that perform these types of tests. The i-STAT CHEM8+ cartridge contains nine measured assays (sodium, potassium, chloride, blood urea nitrogen (BUN), ionized calcium, TCO2, glucose, creatinine and hematocrit) and is categorized as a CLIA waived test for venous whole blood samples, meaning this cartridge could be used in a facility with a CLIA Certificate of Waiver as a waived test.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit