Back to search
FDAFDA · Food and drugs·Recall number Z-1266-2023

Attune Revision Limb Preservation System (LPS) Tibial Insert SM 12MM, Part Number 151760312

Published March 29, 2023 · Recall initiated: February 15, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
DePuy Orthopaedics, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Canada, Poland, Germany, India, Australia, & New Zealand
Lot
GTIN: 10603295490869; Lot: JP6464
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.