Back to search
FDAFDA · Food and drugs·Recall number Z-1274-2013

Zilver® PTX® Drug-Eluting Peripheral Stent - 6 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps Preven the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to:  Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD .  improve blood flow to the extremity.

Published May 22, 2013 · Recall initiated: April 18, 2013

Summary

FDA records medical device Class I recall Z-1274-2013, published May 22, 2013 and initiated April 18, 2013. Recalling firm: Cook, Inc.; Status: Terminated; Reported quantity: 1,057 units in USA; Lot: REF ZIV6-35-125-6-80-PTX ; REFG24890 UDI 10827002248905 All lots distributed prior to 4/16/2013. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cook, Inc.
Distributor
Worldwide Distribution - USA Nationwide; in the countries of Japan, Europe, Brazil, and Australia
Lot
REF ZIV6-35-125-6-80-PTX ; REFG24890 UDI 10827002248905 All lots distributed prior to 4/16/2013
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Zilver® PTX® Drug-Eluting Peripheral Stent - 6 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps Preven the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to:  Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD .  improve blood flow to the extremity. recalled?

Yes. FDA published recall Z-1274-2013 on May 22, 2013. Reported hazard: Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation

What should I do if I have this product?

Voluntary: Firm initiated · Letter