Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX.
Published March 25, 2015 · Recall initiated: February 17, 2015
Summary
FDA records medical device Class II recall Z-1281-2015, published March 25, 2015 and initiated February 17, 2015. Recalling firm: Synthes, Inc.; Status: Terminated; Reported quantity: 183; Lot: Part numbers:03.010.048 and 03.010.052 with Lot Numbers: 4874015 4874021 4927262 4954481 4963192 4874016 4874022 4944750 4957586 4963193 4784017 4899564 4944751 4957587 4972003 4874018 4899565 4954478 4957588 4972004 4874019 4927260 4954479 4957589 4874020 4927261 4954480 4963190 4922533 4922975 4981158 4922970 4923015 4981159 4922971 4977241 4981160 4922972 4977242 4981162 4922973 4981156 5035905 4922974 4981157 5035906. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes, Inc.
- Distributor
- Nationwide Distribution
- Lot
- Part numbers:03.010.048 and 03.010.052 with Lot Numbers: 4874015 4874021 4927262 4954481 4963192 4874016 4874022 4944750 4957586 4963193 4784017 4899564 4944751 4957587 4972003 4874018 4899565 4954478 4957588 4972004 4874019 4927260 4954479 4957589 4874020 4927261 4954480 4963190 4922533 4922975 4981158 4922970 4923015 4981159 4922971 4977241 4981160 4922972 4977242 4981162 4922973 4981156 5035905 4922974 4981157 5035906
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile TubeFebruary 5, 2020
- FDA2.4MM TI VA LOCKING SCREW STARDRIVE 18MM STERILE, Part Number 04.210.118TS - Product Usage:Sterile TubeFebruary 5, 2020
- FDA3.7mm Cannulated Locking Screws and 3.7mm Cannulated Conical Screws; . the Cannulated Screws are indicated forJune 24, 2015
- FDAPull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling isJune 24, 2015
- FDAMi Pepito Camaron Molido Ground Shrimp, Net wt. .75 oz (16 gr), Packaged in a plastic bag, Packed &March 25, 2015
- FDALeMaitre Vascular Tumescent Catheter Inversion (TCI) System. The TCI System is intended to provide a system toMarch 25, 2015
Questions about this recall
Was Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX. recalled?
Yes. FDA published recall Z-1281-2015 on March 25, 2015. Reported hazard: For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.
What should I do if I have this product?
Voluntary: Firm initiated · Letter