Helitene Absorbable Collagen Hemostatic Sponge Helitene is indicated in surgical procedures (other than ophthalmological and urological surgery) as an adjunct to hemostasis when control of bleeding by ligature or conventional procedures is ineffective or impractical.
Published May 22, 2013 · Recall initiated: March 28, 2013
Summary
FDA records medical device Class II recall Z-1290-2013, published May 22, 2013 and initiated March 28, 2013. Recalling firm: Integra LifeSciences Corporation; Status: Terminated; Reported quantity: 14, 525 boxes; Lot: 1110788, 1112217 1110222, 1111357 1111602, 1105255 1110224, 1110225 1111601, 1110223 1110790, 1111603. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Integra LifeSciences Corporation
- Distributor
- Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Britain, Italy, Japan, Peru, Singapore, Thailand, and South Africa.
- Lot
- 1110788, 1112217 1110222, 1111357 1111602, 1105255 1110224, 1110225 1111601, 1110223 1110790, 1111603
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntegra Meshed Dermal Regeneration Template 5 cmx 5 cm (2 in x 2in) Rx Only Meshed Integra® DermalSeptember 21, 2016
- FDAIntegra Newdeal SURFIX Fixed Angle Locking Screw 3.5mm x 22mm (catalog number 285322S) and 3.5mm x 30mmJune 11, 2014
- FDAIntegra PANTA Arthrodesis Nail System. Intended for use in tibiotalocalcaneal arthrodesis and treatment ofJuly 24, 2013
- FDAOra Plug Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen Wound Dressings forMay 22, 2013
- FDAHelistat Absorbable Hemostatic,Collagen Based Helistat is indicated in surgical procedures (other thanMay 22, 2013
- FDAColla Tape Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen Wound Dressings forMay 22, 2013
Questions about this recall
Was Helitene Absorbable Collagen Hemostatic Sponge Helitene is indicated in surgical procedures (other than ophthalmological and urological surgery) as an adjunct to hemostasis when control of bleeding by ligature or conventional procedures is ineffective or impractical. recalled?
Yes. FDA published recall Z-1290-2013 on May 22, 2013. Reported hazard: Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.
What should I do if I have this product?
Voluntary: Firm initiated · Letter