i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and respiratory based acid-based disturbances.
Published April 2, 2014 · Recall initiated: January 1, 2014
Summary
FDA records medical device Class II recall Z-1306-2014, published April 2, 2014 and initiated January 1, 2014. Recalling firm: Abbott Point Of Care Inc.; Status: Terminated; Reported quantity: 48, 450 US, 1400 OUS (49,850 catridges); Lot: List number 03P78-25 :Lot number N13183. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Point Of Care Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Canada.
- Lot
- List number 03P78-25 :Lot number N13183
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAi-STAT EG6+ cartridge; List Number: 03P77-25;May 6, 2026
- FDAi-STAT EG7+ cartridge; List Number: 03P76-25;May 6, 2026
- FDAi-STAT G3+ cartridge; List Number: 03P78-26;May 6, 2026
- FDAi-STAT EG6+ cartridge. List Number: 03P77-25.September 24, 2025
- FDAFIBERNET Embolic Protection System: Recommended Vessel Diameter 6.0 -7.0 mm, Maximum Guidewire Diameter: 0.014April 2, 2014
- FDAHeartMate II LVAD Pump and Pock Controller Kit; Product Usage: HeartMate II is a mechanical device thatApril 2, 2014
Questions about this recall
Was i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and respiratory based acid-based disturbances. recalled?
Yes. FDA published recall Z-1306-2014 on April 2, 2014. Reported hazard: Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results.
What should I do if I have this product?
Voluntary: Firm initiated · Letter