TDWorkstation version 11.01.A or higher. TDWorkstation Evolution. TDWorkstation receives patient results from a connected instrument and sends them to a Laboratory Information System (LIS).
Published April 9, 2014 · Recall initiated: February 17, 2014
Summary
FDA records medical device Class III recall Z-1310-2014, published April 9, 2014 and initiated February 17, 2014. Recalling firm: Technidata S.A.; Status: Terminated; Reported quantity: 85 units total (1 unit in US); Lot: version 11.01.A or higher. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
If a conflict occurs between a Demography Query and a Result Transmission by the instrument, the TDWorkstation could send an incorrect patient demography to an instrument.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Technidata S.A.
- Distributor
- Distributed in the state of PA.
- Lot
- version 11.01.A or higher
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- France
Related recalls
- FDAArtis Zee Angiographic x-ray systems.April 9, 2014
- FDAZoll Medical X Series, PropaqMD, PropaqM Defibrillator/Pacemaker/Monitor With System Software VersionApril 9, 2014
- FDARaySearch Ray Station, Radiation Therapy Treatment Planning System, Stand-alone softwareApril 9, 2014
- FDAPinnacle3 Radiation Therapy Planning (RTP) System, software version 9.0.April 9, 2014
- FDATDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100).July 24, 2013
- FDATD-Synergy. Laboratory Information System.May 29, 2013
Questions about this recall
Was TDWorkstation version 11.01.A or higher. TDWorkstation Evolution. TDWorkstation receives patient results from a connected instrument and sends them to a Laboratory Information System (LIS). recalled?
Yes. FDA published recall Z-1310-2014 on April 9, 2014. Reported hazard: If a conflict occurs between a Demography Query and a Result Transmission by the instrument, the TDWorkstation could send an incorrect patient demography to an instrument.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail