TDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100). Intended use: Laboratory Information System (L.I.S)
Published July 24, 2013 · Recall initiated: June 20, 2013
Summary
FDA records medical device Class III recall Z-1743-2013, published July 24, 2013 and initiated June 20, 2013. Recalling firm: Technidata S.A.; Status: Terminated; Reported quantity: 1 site; Lot: All TDHistology/Cytology product versions. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A recall was initiated by Technidata because use of symbols (less than or equal to or greater than or equal to or any other symbol) in Microsoft Word reports when using electronic transmissions of results by of HPRIM (including HPRIM Medecin via Kermit) or HL7 must be absolutely prohibited. Such symbols are not supported by the Microsoft Word interface used by the Histology/Cytology software and
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Technidata S.A.
- Distributor
- US Distribution CA only.
- Lot
- All TDHistology/Cytology product versions.
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- France
Related recalls
- FDATDWorkstation version 11.01.A or higher. TDWorkstation Evolution. TDWorkstation receives patient results fromApril 9, 2014
- FDARemB Electric Universal Driver REF 6400-099-000 For use with the CORE and TPS System sRx Only The RemBJuly 24, 2013
- FDAPolarizer D Fixed w/Rotating Red 1 lambda plate. Utilized in upright Zeiss microscopes to aid in diagnosis byJuly 24, 2013
- FDACORE Universal Driver REF 5400- 99 Rx Only This drill may also be used with theTotal Performance SystemJuly 24, 2013
- FDAMulti Vector Distractor Pin Holding Clamp, limited bone stock The holding clamp is part of the Multi VectorJuly 24, 2013
- FDATD-Synergy. Laboratory Information System.May 29, 2013
Questions about this recall
Was TDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100). Intended use: Laboratory Information System (L.I.S) recalled?
Yes. FDA published recall Z-1743-2013 on July 24, 2013. Reported hazard: A recall was initiated by Technidata because use of symbols (less than or equal to or greater than or equal to or any other symbol) in Microsoft Word reports when using electronic transmissions of results by of HPRIM (including HPRIM Medecin via Kermit) or HL7 must be absolutely prohibited. Such symbols are not supported by the Microsoft Word interface used by the Histology/Cytology software and
What should I do if I have this product?
Voluntary: Firm initiated · Letter