Sekisui Diagnostics Control Kit OSOM® BVBLUE® Bacterial Vaginosis
Published March 27, 2024 · Recall initiated: January 9, 2024
Summary
FDA records medical device Class II recall Z-1311-2024, published March 27, 2024 and initiated January 9, 2024. Recalling firm: Mckesson Medical-Surgical Inc. Corporate Office; Status: Ongoing; Reported quantity: 1 unit; Lot: Catalogue Number: 184; Lot No. CK233. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes .
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Mckesson Medical-Surgical Inc. Corporate Office
- Distributor
- US Nationwide distribution in the states of AZ, PA and TX.
- Lot
- Catalogue Number: 184; Lot No. CK233;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPROLIA (denosumab), injection, 60mg/mL, For Subcutaneous Use Only, Rx Only, Manufactured by: Amgen Inc.,July 30, 2025
- FDAEVENITY, (romosozumab -aqqg) injection, 105mg/1.17 mL, For Subcutaneous Use Only, Rx Only, Manufactured by:July 30, 2025
- FDABENLYSTA (belimumab) for injection, 400 mg/20 mL vial, Rx only, NDC 49401-102-01, GSK.July 30, 2025
- FDASEKISUI DIAGNOSTICS LLC SERUM HCG OSOMGENZME CONTROL SET.June 4, 2025
- FDANautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures toMarch 27, 2024
- FDAAbbott Rapid Dx N America LLC Reproductive Endocrinology Control Set McKesson Consult" Pregnancy (hCG) TestingMarch 27, 2024
Questions about this recall
Was Sekisui Diagnostics Control Kit OSOM® BVBLUE® Bacterial Vaginosis recalled?
Yes. FDA published recall Z-1311-2024 on March 27, 2024. Reported hazard: Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes .
What should I do if I have this product?
Voluntary: Firm initiated · Telephone