Back to search
FDAFDA · Food and drugs·Recall number Z-1312-2024

Abbott Rapid Dx N America LLC Reproductive Endocrinology Control Set McKesson Consult" Pregnancy (hCG) Testing Positive Level / Negative Level 2 X 4 mL

Published March 27, 2024 · Recall initiated: January 9, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.

What to do

Voluntary: Firm initiated · Telephone

Recall details

Manufacturer
Mckesson Medical-Surgical Inc. Corporate Office
Distributor
US Nationwide distribution in the states of AZ, PA and TX.
Lot
Catalogue Number: 5011; UDI/DI: 612479202788; Lot No. KN00589;
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.