MINI STICK MAX 4F X 10 CM STD .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-750
Published March 27, 2024 · Recall initiated: January 30, 2024
Summary
FDA records medical device Class II recall Z-1335-2024, published March 27, 2024 and initiated January 30, 2024. Recalling firm: Angiodynamics, Inc.; Status: Ongoing; Reported quantity: 4366 units; Lot: UDI: 15051684022934 UPN: H965457501 Lot Number: 5784035 5784036 5784136 5787837 5788554 5788555 5789978 5790448 5790449 5791589 5791590 5791591 5797460 5797461 5798838 5799613 5799623 5800399 5804390 5806116. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Angiodynamics, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.
- Lot
- UDI: 15051684022934 UPN: H965457501 Lot Number: 5784035 5784036 5784136 5787837 5788554 5788555 5789978 5790448 5790449 5791589 5791590 5791591 5797460 5797461 5798838 5799613 5799623 5800399 5804390 5806116
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAccuVu Angiographic Catheter, Straight Flush, Non-Braided, With 10 R/O Markers; ACCU-VU STR FL 4F X 100CM 035April 29, 2026
- FDASoft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 65CM 038 NB 0SH; Catalog No.: 10714014;April 29, 2026
- FDASOFT-VU KUMPE 5F X 65CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732701August 7, 2024
- FDASOFT-VU KUMPE 5F X 40CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732702August 7, 2024
- FDAMINI STICK MAX 4F X 10 CM STD .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewireMarch 27, 2024
- FDAMINI STICK MAX 4F X 10 CM STIFF .018 SS/SS ECHO 2.75" PG-Used for the percutaneous introduction of a guidewireMarch 27, 2024
Questions about this recall
Was MINI STICK MAX 4F X 10 CM STD .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-750 recalled?
Yes. FDA published recall Z-1335-2024 on March 27, 2024. Reported hazard: Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
What should I do if I have this product?
Voluntary: Firm initiated · Letter