Brilliance CT 16-Slice Water, System Code: 728245; Brilliance CT 10-Slice Water, System Code: 728250; Brilliance CT 6-Slice Water, System Code: 728255; MX8000 IDT 16, System Code: 728120 & MX8000 IDT 10, System Code: 728125 Computed Tomography X-ray systems. Distributed by Philips Medical Systems (Cleveland), Inc., 595 Miner Road, Cleveland, OH 44143 Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Published May 29, 2013 · Recall initiated: March 12, 2013
Summary
FDA records medical device Class II recall Z-1338-2013, published May 29, 2013 and initiated March 12, 2013. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 515 units; Lot: System Code #728255, Serial #: 9702, 9703, 9704, 9705, 9706, 9707, 9709, 9710, 9713, 9714, 9717, 9718, 9719, 9721, 9722, 9723, 9724, 9725, 9727, 9730, 9731, 9732, 9733, 9734, 9735, 9738, 9739, 9740, 9741, 9742, 9743, 9746, 9747, 9749, 9752, 9753, 9756, 9758, 9759, 9760, 9761, 9762, 9765, 9766, 9768, 9769, 9770, 9772, 9774, 9778, 9780, 9781, 9783, 9784, 9786, 9787, 9790, 9791, 9792, 9793, 9796, 9797, 9801, 9805, 9807, 9808 & 9810. System Code #728250, Serial #: 2556, 2557, 2560, 2562, 2563, 2565, 2567, 2568, 2569, 2570, 2571, 2574, 2576, 2577, 2579, 2580, 2581, 2584, 2585, 2587, 2588, 2590, 2591, 2592, 2593, 2594, 2595, 2597, 9738 & 9741. System Code #728245, Serial #: 2562, 2564, 2567, 2573, 2576, 2582, 2584, 2585, 2586, 2587, 2589, 2590, 2594, 2596, 2597, 3014, 3091, 3105, 3151, 3217, 3235, 3287, 3303, 3312, 3339, 3340, 3341, 3342, 3343, 3344, 3346, 3348, 3351, 3353, 3355, 3356, 3357, 3358, 3359, 3360, 3363, 3364, 3366, 3369, 3371, 3373, 3374, 3377, 3378, 3379, 3380, 3381, 3382, 3383, 3385, 3388, 3392, 3393, 3394, 3397, 3400, 3404, 3406, 3407, 3408, 3410, 3413, 3414, 3415, 3416, 3419, 3423, 3425, 3427, 3429, 3431, 3432, 3433, 3435, 3436, 3437, 3438, 3439, 3440, 3441, 3443, 3445, 3446, 3448, 3449, 3451, 3452, 3453, 3454, 3455, 3458, 3459, 3461, 3462, 3464, 3468, 3469, 3470, 3471, 3473, 3474, 3475, 3477, 3478, 3479, 3481, 3482, 3483, 3484, 3485, 3486, 3488, 3489, 3491, 3492, 3493, 3495, 3496, 3497, 3498, 3499, 3500, 3501, 3502, 3503, 3504, 3505, 3508, 3509, 3510, 3511, 3512, 3513, 3514, 3515, 3517, 9716, 9718, 9719, 9724, 9732, 9741, 9743, 9746, 9749, 9778, 9780, 9787, 9790, 9797, 9801, 9808 & 9810. System Code #728120, Serial #: 2502, 2510, 2513, 2516, 2517, 2518, 2525, 2528, 2531, 2533, 2535, 2539, 2546, 3006, 3008, 3013, 3014, 3021, 3025, 3026, 3027, 3028, 3029, 3030, 3035, 3037, 3039, 3040, 3043, 3044, 3047, 3049, 3050, 3051, 3053, 3054, 3056, 3059, 3061, 3062, 3063, 3064, 3068, 3069, 3070, 3072, 3073, 3075, 3076, 3077, 3078, 3079, 3081, 3082, 3084, 3085, 3087, 3088, 3089, 3091, 3093, 3094, 3095, 3097, 3099, 3100, 3103, 3104, 3105, 3106, 3107, 3108, 3110, 3111, 3112, 3113, 3114, 3116, 3117, 3118, 3119, 3120, 3121, 3122, 3124, 3125, 3126, 3127, 3129, 3130, 3131, 3132, 3135, 3136, 3137, 3138, 3141, 3143, 3144, 3146, 3148, 3149, 3150, 3151, 3152, 3153, 3154, 3155, 3156, 3157, 3159, 3160, 3161, 3162, 3163, 3164, 3165, 3167, 3168, 3169, 3171, 3172, 3174, 3176, 3177, 3178, 3179, 3180, 3182, 3183, 3185, 3187, 3188, 3189, 3190, 3192, 3194, 3195, 3196, 3198, 3199, 3202, 3203, 3204, 3205, 3206, 3207, 3208, 3209, 3211, 3212, 3213, 3214, 3216, 3217, 3218, 3219, 3220, 3221, 3222, 3223, 3224, 3225, 3229, 3230, 3232, 3233, 3234, 3235, 3236, 3237, 3238, 3239, 3240, 3245, 3246, 3247, 3248, 3250, 3251, 3252, 3253, 3254, 3255, 3258, 3261, 3262, 3264, 3265, 3266, 3268, 3269, 3270, 3271, 3273, 3274, 3277, 3279, 3280, 3281, 3283, 3284, 3285, 3287, 3289, 3290, 3293, 3294, 3295, 3296, 3297, 3298, 3299, 3300, 3302, 3303, 3304, 3305, 3307, 3308, 3310, 3311, 3312, 3315, 3316, 3318, 3319, 3320, 3321, 3322, 3324, 3328, 3329, 3330, 3367 & 9760. System Code #728125, Serial #: 2501, 2505, 2508, 2509, 2513, 2515, 2516, 2522, 2523, 2524, 2525, 2526, 2529, 2530, 2531, 2533, 2534, 2535, 2539, 2540, 2542, 2543, 2548 & 2549. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Philips Healthcare received a report from the field stating when they viewed the CTDIVol information from the PACS system, the calculations were too high. Instead of storing the CTDIVol value for each slice in the DICOM tag, the system took the sum of all CTDIVol values from all the images in the dataset and stored this value in the DICOM tag, resulting in incorrect information.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Worldwide distribution: USA (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, VA, VT, WA, WI, WV and WY; countries of: Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Denmark, France, Germany, Gibraltar, India, Iraq, Ireland, Israel, Italy, Japan, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Philippines, Poland, Portugal, Singapore, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Venezuela.
- Lot
- System Code #728255, Serial #: 9702, 9703, 9704, 9705, 9706, 9707, 9709, 9710, 9713, 9714, 9717, 9718, 9719, 9721, 9722, 9723, 9724, 9725, 9727, 9730, 9731, 9732, 9733, 9734, 9735, 9738, 9739, 9740, 9741, 9742, 9743, 9746, 9747, 9749, 9752, 9753, 9756, 9758, 9759, 9760, 9761, 9762, 9765, 9766, 9768, 9769, 9770, 9772, 9774, 9778, 9780, 9781, 9783, 9784, 9786, 9787, 9790, 9791, 9792, 9793, 9796, 9797, 9801, 9805, 9807, 9808 & 9810. System Code #728250, Serial #: 2556, 2557, 2560, 2562, 2563, 2565, 2567, 2568, 2569, 2570, 2571, 2574, 2576, 2577, 2579, 2580, 2581, 2584, 2585, 2587, 2588, 2590, 2591, 2592, 2593, 2594, 2595, 2597, 9738 & 9741. System Code #728245, Serial #: 2562, 2564, 2567, 2573, 2576, 2582, 2584, 2585, 2586, 2587, 2589, 2590, 2594, 2596, 2597, 3014, 3091, 3105, 3151, 3217, 3235, 3287, 3303, 3312, 3339, 3340, 3341, 3342, 3343, 3344, 3346, 3348, 3351, 3353, 3355, 3356, 3357, 3358, 3359, 3360, 3363, 3364, 3366, 3369, 3371, 3373, 3374, 3377, 3378, 3379, 3380, 3381, 3382, 3383, 3385, 3388, 3392, 3393, 3394, 3397, 3400, 3404, 3406, 3407, 3408, 3410, 3413, 3414, 3415, 3416, 3419, 3423, 3425, 3427, 3429, 3431, 3432, 3433, 3435, 3436, 3437, 3438, 3439, 3440, 3441, 3443, 3445, 3446, 3448, 3449, 3451, 3452, 3453, 3454, 3455, 3458, 3459, 3461, 3462, 3464, 3468, 3469, 3470, 3471, 3473, 3474, 3475, 3477, 3478, 3479, 3481, 3482, 3483, 3484, 3485, 3486, 3488, 3489, 3491, 3492, 3493, 3495, 3496, 3497, 3498, 3499, 3500, 3501, 3502, 3503, 3504, 3505, 3508, 3509, 3510, 3511, 3512, 3513, 3514, 3515, 3517, 9716, 9718, 9719, 9724, 9732, 9741, 9743, 9746, 9749, 9778, 9780, 9787, 9790, 9797, 9801, 9808 & 9810. System Code #728120, Serial #: 2502, 2510, 2513, 2516, 2517, 2518, 2525, 2528, 2531, 2533, 2535, 2539, 2546, 3006, 3008, 3013, 3014, 3021, 3025, 3026, 3027, 3028, 3029, 3030, 3035, 3037, 3039, 3040, 3043, 3044, 3047, 3049, 3050, 3051, 3053, 3054, 3056, 3059, 3061, 3062, 3063, 3064, 3068, 3069, 3070, 3072, 3073, 3075, 3076, 3077, 3078, 3079, 3081, 3082, 3084, 3085, 3087, 3088, 3089, 3091, 3093, 3094, 3095, 3097, 3099, 3100, 3103, 3104, 3105, 3106, 3107, 3108, 3110, 3111, 3112, 3113, 3114, 3116, 3117, 3118, 3119, 3120, 3121, 3122, 3124, 3125, 3126, 3127, 3129, 3130, 3131, 3132, 3135, 3136, 3137, 3138, 3141, 3143, 3144, 3146, 3148, 3149, 3150, 3151, 3152, 3153, 3154, 3155, 3156, 3157, 3159, 3160, 3161, 3162, 3163, 3164, 3165, 3167, 3168, 3169, 3171, 3172, 3174, 3176, 3177, 3178, 3179, 3180, 3182, 3183, 3185, 3187, 3188, 3189, 3190, 3192, 3194, 3195, 3196, 3198, 3199, 3202, 3203, 3204, 3205, 3206, 3207, 3208, 3209, 3211, 3212, 3213, 3214, 3216, 3217, 3218, 3219, 3220, 3221, 3222, 3223, 3224, 3225, 3229, 3230, 3232, 3233, 3234, 3235, 3236, 3237, 3238, 3239, 3240, 3245, 3246, 3247, 3248, 3250, 3251, 3252, 3253, 3254, 3255, 3258, 3261, 3262, 3264, 3265, 3266, 3268, 3269, 3270, 3271, 3273, 3274, 3277, 3279, 3280, 3281, 3283, 3284, 3285, 3287, 3289, 3290, 3293, 3294, 3295, 3296, 3297, 3298, 3299, 3300, 3302, 3303, 3304, 3305, 3307, 3308, 3310, 3311, 3312, 3315, 3316, 3318, 3319, 3320, 3321, 3322, 3324, 3328, 3329, 3330, 3367 & 9760. System Code #728125, Serial #: 2501, 2505, 2508, 2509, 2513, 2515, 2516, 2522, 2523, 2524, 2525, 2526, 2529, 2530, 2531, 2533, 2534, 2535, 2539, 2540, 2542, 2543, 2548 & 2549.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDAVITROS Immunodiagnostic Products CK-MB Calibrators, REF 128 6293, IVD, Ortho-Clinical Diagnostics High WycombeMay 29, 2013
- FDALATITUDE® Patient Management System, Model 6488, Version 7.3. The LATITUDE Patient Management system isMay 29, 2013
Questions about this recall
Was Brilliance CT 16-Slice Water, System Code: 728245; Brilliance CT 10-Slice Water, System Code: 728250; Brilliance CT 6-Slice Water, System Code: 728255; MX8000 IDT 16, System Code: 728120 & MX8000 IDT 10, System Code: 728125 Computed Tomography X-ray systems. Distributed by Philips Medical Systems (Cleveland), Inc., 595 Miner Road, Cleveland, OH 44143 Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories. recalled?
Yes. FDA published recall Z-1338-2013 on May 29, 2013. Reported hazard: Philips Healthcare received a report from the field stating when they viewed the CTDIVol information from the PACS system, the calculations were too high. Instead of storing the CTDIVol value for each slice in the DICOM tag, the system took the sum of all CTDIVol values from all the images in the dataset and stored this value in the DICOM tag, resulting in incorrect information.
What should I do if I have this product?
Voluntary: Firm initiated · Letter