FDAFDA · Food and drugs·Recall number Z-1340-2013

Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical endoscopic camera system during surgical procedures.

Published May 29, 2013 · Recall initiated: March 18, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key strokes (Caps lock and Num Lock) which may lock up and requires the user to reboot to continue.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Linvatec Corp. dba ConMed Linvatec
Distributor
Worldwide Distribution-USA (nationwide) including the Virgin Islands and the states of CA, CT, FL, GA, IN, KY, KS, LA, MD, MI, MN, MO, MS, MT, NE, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, and WI, and the countries of Austria, Belgium, Canada, Denmark, Hong Kong and Zambia.
Lot
VP1600
Risk classification
Class II
Area (FDA)
device
Country
United States