Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical endoscopic camera system during surgical procedures.
Published May 29, 2013 · Recall initiated: March 18, 2013
Summary
FDA records medical device Class II recall Z-1340-2013, published May 29, 2013 and initiated March 18, 2013. Recalling firm: Linvatec Corp. dba ConMed Linvatec; Status: Terminated; Reported quantity: 267 units; Lot: VP1600. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key strokes (Caps lock and Num Lock) which may lock up and requires the user to reboot to continue.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Linvatec Corp. dba ConMed Linvatec
- Distributor
- Worldwide Distribution-USA (nationwide) including the Virgin Islands and the states of CA, CT, FL, GA, IN, KY, KS, LA, MD, MI, MN, MO, MS, MT, NE, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, and WI, and the countries of Austria, Belgium, Canada, Denmark, Hong Kong and Zambia.
- Lot
- VP1600
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA***REF C7120***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention duringJuly 17, 2013
- FDA***REF C7122***QTY 1***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissentionJuly 17, 2013
- FDAGENESYS Cross FT Suture Anchor with Two #2 (5 metric) Hi-FI Registered Trade Mark Sutures Rx Only STERILE EOJuly 10, 2013
- FDALATITUDE® Patient Management System, Model 6488, Version 7.3. The LATITUDE Patient Management system isMay 29, 2013
- FDARostorelle L, Polypropylene Mesh, Catalog number 501440, Manufacturer Coloplast A/S 3050 Denmark.May 29, 2013
- FDAD3000 Advantage Drive System, 115v Product Usage: The Advantage Drive Systems function as a powered instrumentMarch 13, 2013
Questions about this recall
Was Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical endoscopic camera system during surgical procedures. recalled?
Yes. FDA published recall Z-1340-2013 on May 29, 2013. Reported hazard: The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key strokes (Caps lock and Num Lock) which may lock up and requires the user to reboot to continue.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit