FDAFDA · Food and drugs·Recall number Z-1680-2013

***REF C7120***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthroscopic procedures.

Published July 17, 2013 · Recall initiated: May 30, 2013

Summary

FDA records medical device Class II recall Z-1680-2013, published July 17, 2013 and initiated May 30, 2013. Recalling firm: Linvatec Corp. dba ConMed Linvatec; Status: Terminated; Reported quantity: 4,691 Units total of each; Lot: C7120 Lot's #: 1301184, 1301214, 1301254, 1301284, 1301294, 1302014, 1302044, 1302054, 1302064, 1302074, 13020124, 1302134, 1302144, 1302194, 1302204, 1302214, 1302224, 1302254, 1302264, 1303014, 1303054, 1303064, 1303074, 1303114, 1303124, 1303204, 1303214, 1303264, 1303274, 1303284, 1304014, 1304034, 1304054, 1304154, 1304294, 1304304, and 1305014. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Linvatec Corp. dba ConMed Linvatec
Distributor
Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, GA, IL, IN, KS, MO, NC, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, and WI, and the countries of Austria; Belgium, Canada, China, Czech Republic, Finland, France, Great Britain, Italy, and Japan.
Lot
C7120 Lot's #: 1301184, 1301214, 1301254, 1301284, 1301294, 1302014, 1302044, 1302054, 1302064, 1302074, 13020124, 1302134, 1302144, 1302194, 1302204, 1302214, 1302224, 1302254, 1302264, 1303014, 1303054, 1303064, 1303074, 1303114, 1303124, 1303204, 1303214, 1303264, 1303274, 1303284, 1304014, 1304034, 1304054, 1304154, 1304294, 1304304, and 1305014.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ***REF C7120***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthroscopic procedures. recalled?

Yes. FDA published recall Z-1680-2013 on July 17, 2013. Reported hazard: Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit