MINI STICK MAX 5F X 10 CM STIFF .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-759

Published March 27, 2024 · Recall initiated: January 30, 2024

Summary

FDA records medical device Class II recall Z-1341-2024, published March 27, 2024 and initiated January 30, 2024. Recalling firm: Angiodynamics, Inc.; Status: Ongoing; Reported quantity: 5840 units; Lot: UDI: 15051684023023 UPN: H965457591 Lot Number: 5789493 5789494 5789495 5789496 5789497 5791596 5791597 5791598 5793392 5793393 5794602 5794603 5794604 5794605 5794606 5795874 5795875 5797457 5797458 5797466 5797467 5798837 5798843 5798844. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Angiodynamics, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.
Lot
UDI: 15051684023023 UPN: H965457591 Lot Number: 5789493 5789494 5789495 5789496 5789497 5791596 5791597 5791598 5793392 5793393 5794602 5794603 5794604 5794605 5794606 5795874 5795875 5797457 5797458 5797466 5797467 5798837 5798843 5798844
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was MINI STICK MAX 5F X 10 CM STIFF .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-759 recalled?

Yes. FDA published recall Z-1341-2024 on March 27, 2024. Reported hazard: Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

What should I do if I have this product?

Voluntary: Firm initiated · Letter