Back to search
FDAFDA · Food and drugs·Recall number Z-1343-2024

MINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-765

Published March 27, 2024 · Recall initiated: January 30, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Angiodynamics, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.
Lot
UDI: 15051684023085 UPN: H965457651 Lot Number: 5790681 5795961 5800479 5803518 5805067
Risk classification
Class II
Area (FDA)
device
Country
United States