SIGNA Pioneer MR Systems Product Usage: The SIGNA Pioneer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA Pioneer reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Published April 13, 2016 · Recall initiated: March 1, 2016
Summary
FDA records medical device Class II recall Z-1347-2016, published April 13, 2016 and initiated March 1, 2016. Recalling firm: GE Medical Systems, LLC; Status: Terminated; Reported quantity: 16 (2 US; 14 OUS); Lot: Mfg Lot or Serial # System ID 001612TA1 N/A 001718TA6 336397P3T 002331TA7 083132425984 001719TA4 250598MR1 001424TA1 M2850013 001537TA0 TP0008MR07 001662TA6 157022MR01 001184TA1 ZC0002 001652TA7 N/A 002227TA7 ZC0003 002278TA0 ZC0004 002330TA9 ZC0006 002529TA6 N/A 001485TA2 34026MRS05 002279TA8 34225MRS02 002395TA2 34185MRS02. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ferrous material in the SIGNA Pioneer Table Control Box (TCB) could cause unexpected attraction to the MR system's high strength magnetic field during servicing activities. This could create a potential risk of serious injury to service personnel who attempt to inappropriately remove the TCB from the magnetic field. There have been no events reported as a result of this issue. Clinical scanni
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Medical Systems, LLC
- Distributor
- Worldwide Distribution - US Nationwide in the states of SC, NC and the countries of Australia, Canada, France. Germany, Japan and Turkey.
- Lot
- Mfg Lot or Serial # System ID 001612TA1 N/A 001718TA6 336397P3T 002331TA7 083132425984 001719TA4 250598MR1 001424TA1 M2850013 001537TA0 TP0008MR07 001662TA6 157022MR01 001184TA1 ZC0002 001652TA7 N/A 002227TA7 ZC0003 002278TA0 ZC0004 002330TA9 ZC0006 002529TA6 N/A 001485TA2 34026MRS05 002279TA8 34225MRS02 002395TA2 34185MRS02
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASenographe Pristina 3DNovember 11, 2026
- FDASenographe PristinaNovember 11, 2026
- FDASenographe PristinaNovember 11, 2026
- FDAOptima XR240amx, Mobile Digital Radiographic SystemJuly 22, 2026
- FDANAMIC CONVENIENCE KIT MERCY GENERAL HOSPITAL NEURO KIT, REF Catalog No. 601319810, UPN Product No.April 13, 2016
- FDAVITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & 10758750007110. Intended for useApril 13, 2016
Questions about this recall
Was SIGNA Pioneer MR Systems Product Usage: The SIGNA Pioneer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA Pioneer reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. recalled?
Yes. FDA published recall Z-1347-2016 on April 13, 2016. Reported hazard: Ferrous material in the SIGNA Pioneer Table Control Box (TCB) could cause unexpected attraction to the MR system's high strength magnetic field during servicing activities. This could create a potential risk of serious injury to service personnel who attempt to inappropriately remove the TCB from the magnetic field. There have been no events reported as a result of this issue. Clinical scanni
What should I do if I have this product?
Voluntary: Firm initiated · Letter