Back to search
FDAFDA · Food and drugs·Recall number Z-1356-2020

Zimmer Sterile Disposable Tourniquet Cuff with Protective Sleeve and PLC, Dual Port, Dual Bladder, Item No. 60708015200, UDI (01)00889024377547 (17)230812(10)27968701 The product is a tourniquet - Product Usage: Tourniquets are intended to be used by qualified medical professionals to exert enough pressure on the arterial blood flow in a limb to produce a bloodless operating field.

Published March 4, 2020 · Recall initiated: January 3, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When inflating/deflating a single bladder, both bladders could potentially inflate/deflate due to a leak between the bladders.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Zimmer Surgical Inc
Distributor
Worldwide distribution - US Nationwide distribution in the states of CA, FL, IA, ID, IL, LA, MA, MO, NC, NY, OH, OR, PA, SC, TN, TX, WA, WI, and WV and country of Netherlands.
Lot
Lot 27968701
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card