Protective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory (i.e. protective cover) for the Vivo 50 and Vivo 65 ventilators - Product Usage: The protective cover is intended for additional protection of the Vivo 50, Vivo 55, Vivo 60 and Vivo 65 ventilators during transportation, and in hospital, institutional, and home care environments. It can be used while the Vivo ventilators is operating, for example mounted on a wheelchair, in a personal vehicle, or carried by hand. The device is not to be used with the protective cover while positioned in a warm place, such as in direct sunlight or close to a radiator. The protective cover includes the following features: transparent window for accessing front panel and buttons; port for patient circuit, cables, O2 inlet, and standby button; cooling air inlet; patient air inlet; straps for safe mounting; and cooling air outlet.
Published March 4, 2020 · Recall initiated: January 28, 2020
Summary
FDA records medical device Class II recall Z-1352-2020, published March 4, 2020 and initiated January 28, 2020. Recalling firm: Breas Medical, Inc.; Status: Terminated; Reported quantity: 629; Lot: The cover does not have a lot number; these covers were distributed between 01/11/2018 and 12/17/2019. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The outlet hole of the exhalation valve control pressure tube on the ventilator may become obstructed by the protective cover.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Breas Medical, Inc.
- Distributor
- Worldwide distribution - US Nationwide distribution in the states of AR, CA, FL, GA, IA, IL, KS, MA, MI, MO, NJ, NY, OH, PR, SC, TX, VA, WI and countries distribution worldwide.
- Lot
- The cover does not have a lot number; these covers were distributed between 01/11/2018 and 12/17/2019
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVivo 45 LS, pressure and volume ventilator capable of delivering continuous or intermittent ventilatorySeptember 11, 2024
- FDAEllipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36CMarch 4, 2020
- FDAZimmer Sterile Disposable Tourniquet Cuff with Protective Sleeve and PLC, Dual Port, Dual Bladder, Item No.March 4, 2020
- FDATHP Hip Plating System, Item Nos. 14-451070 14-451075 14-451080 14-451085 14-451090 14-451095 14-451100March 4, 2020
- FDAMedline ROBOTIC-LF REF CDS984543B Lot 20AKA927 Expiration 2020-09-30 GTIN (01)10888277056350March 4, 2020
- FDAPhilips DigiTrak XT Holter Recorder Software version number B.1 and C.2. The DigiTrak XT Recorder is availableMarch 4, 2020
Questions about this recall
Was Protective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory (i.e. protective cover) for the Vivo 50 and Vivo 65 ventilators - Product Usage: The protective cover is intended for additional protection of the Vivo 50, Vivo 55, Vivo 60 and Vivo 65 ventilators during transportation, and in hospital, institutional, and home care environments. It can be used while the Vivo ventilators is operating, for example mounted on a wheelchair, in a personal vehicle, or carried by hand. The device is not to be used with the protective cover while positioned in a warm place, such as in direct sunlight or close to a radiator. The protective cover includes the following features: transparent window for accessing front panel and buttons; port for patient circuit, cables, O2 inlet, and standby button; cooling air inlet; patient air inlet; straps for safe mounting; and cooling air outlet. recalled?
Yes. FDA published recall Z-1352-2020 on March 4, 2020. Reported hazard: The outlet hole of the exhalation valve control pressure tube on the ventilator may become obstructed by the protective cover.
What should I do if I have this product?
Voluntary: Firm initiated · Letter