Back to search
FDAFDA · Food and drugs·Recall number Z-1366-2022

Wright EVOLVE TRIAD Plate Cutter, REF 49510120

Published July 20, 2022 · Recall initiated: June 14, 2022

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Wright Medical Technology, Inc.
Distributor
Nationwide and PR, and Australia, Canada, Chile, Colombia, Hong Kong, Puerto Rico
Lot
Lot codes: 955401, 1168853, 1349302, 1956936, 2486857
Risk classification
Class II
Area (FDA)
device
Country
United States