Wright EVOLVE TRIAD Plate Cutter, REF 49510120

Published July 20, 2022 · Recall initiated: June 14, 2022

Summary

FDA records medical device Class II recall Z-1366-2022, published July 20, 2022 and initiated June 14, 2022. Recalling firm: Wright Medical Technology, Inc.; Status: Ongoing; Reported quantity: 231; Lot: Lot codes: 955401, 1168853, 1349302, 1956936, 2486857. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Wright Medical Technology, Inc.
Distributor
Nationwide and PR, and Australia, Canada, Chile, Colombia, Hong Kong, Puerto Rico
Lot
Lot codes: 955401, 1168853, 1349302, 1956936, 2486857
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Wright EVOLVE TRIAD Plate Cutter, REF 49510120 recalled?

Yes. FDA published recall Z-1366-2022 on July 20, 2022. Reported hazard: The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use

What should I do if I have this product?

Voluntary: Firm initiated · Letter