Stryker Infinity Alignment Frame Distal Sub-Assembly, Catalog #33600020, non-sterile.
Published July 12, 2023 · Recall initiated: June 13, 2023
Summary
FDA records medical device Class II recall Z-2105-2023, published July 12, 2023 and initiated June 13, 2023. Recalling firm: Wright Medical Technology, Inc.; Status: Ongoing; Reported quantity: 106 devices; Lot: Lot numbers 2656815 and 2762126, UDI-DI 00889797003926. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventing pin placement through the pin holes.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Wright Medical Technology, Inc.
- Distributor
- There was no U.S. distribution, including government and military. Foreign distribution was made to Australia, Belgium, Canada, Chile, China, Colombia, France, Germany, Italy, Mexico, Netherlands, Spain, Sweden, Switzerland, Taiwan, and the United Kingdom.
- Lot
- Lot numbers 2656815 and 2762126, UDI-DI 00889797003926.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating SystemOctober 7, 2026
- FDAstryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplastyApril 10, 2024
- FDAPANCREAZE (pancrelipase) Delayed-Release Capsules, 100-count bottles, Rx only, Rx only, Manufactured by VIVUSJuly 19, 2023
- FDASurgiMend PRS Meshed; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Size 10cm x 20cmJuly 12, 2023
- FDAStryker SWANSON Flexspan Finger Joint Implant W/O Grommets Silicone, Size: 00, REF 4700020, Sterile.August 31, 2022
- FDAWright EVOLVE TRIAD Plate Cutter, REF 49510120July 20, 2022
Questions about this recall
Was Stryker Infinity Alignment Frame Distal Sub-Assembly, Catalog #33600020, non-sterile. recalled?
Yes. FDA published recall Z-2105-2023 on July 12, 2023. Reported hazard: The devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventing pin placement through the pin holes.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail