MicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, Helical, Sterile, Rx, REF numbers 100101H2HS-V; 100102H2HS-V, 100103H2HS-V, 100104H2HS-V, 100151H2HS-V; 100152H2HS-V; 100153H2HS-V; 100154H2HS-V; 100201H2HS-V; 100202H2HS-V; 100203H2HS-V; 100204H2HS-V; 100206H2HS-V, 100208H2HS-V, 100254H2HS-V, 100256H2HS-V, 100304H2HS-V; 100306H2HS-V; 100308H2HS-V; 100310H2HS-V; 100404H2HS-V; 100406H2HS-V, 100408H2HS-V, 100410H2HS-V, 100506H2HS-V, 100508H2HS-V; 100510H2HS-V, 100515H2HS-V, 100520H2HS-V, 100620H2HS-V, and MV-00208HHSA - Product Usage: The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.

Published March 4, 2020 · Recall initiated: November 22, 2019

Summary

FDA records medical device Class II recall Z-1392-2020, published March 4, 2020 and initiated November 22, 2019. Recalling firm: Microvention, Inc.; Status: Terminated; Reported quantity: 2,762 units; Lot: Catalog/lot numbers: 100101H2HS-V - 1903135UR, 1903185UR, 1903225UR, and 1904035UR; 100102H2HS - V -1903135UR, 1903155UR, 1903155ZR, 1903185UR, 1903205UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR,1904155UR, 1904175UR, 1904175ZR, and 1904245UR; 100103H2HS-V - 1903015UR, 1903045UR, 1903065UR, 1903085UR, 1903135UR, 1903205UR 1904055UR, 1904085UR, 1904155UR, 1905085UR, 1905105UR, 1905135UR, and 1905155UR; 100104H2HS-V - 1903015UR, 1903135UR, 1903205UR, 1904085UR, 1904105UR, and 1904155UR; 100151H2HS-V - 1903085UR, 1903185UR, and 1904015UR; 100152H2HS-V - 1903135UR, 1903155UR, 1903205NR, 1903205UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR, 1904155UR, 1904175UR, 1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR, 1905035UR, 1905065UR, 1905085UR, 1903045WR, 1903185WR, 1903205WR, 1904105WR, 1904175WR, 1904225WR, 1905065WR, and 1905175WR; 100153H2HS-V - 1903015UR, 1903065UR, 1903085UR, 1903185UR, 1904015UR, 1904015ZR, 1904105UR, 1904155UR, 1904245UR, 1904265UR, and 1904295UR; 100154H2HS-V - 1903135UR, 1903155UR, 1903205UR, 1903225UR, 1904015UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR, 1904155UR, 1904175UR, 1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR, 1905035UR, 1905065NR, 1905065UR, 1905155NR, 1903185WR, 1903205WR, 1904105WR, 1904225WR, and 1905065WR; 100201H2HS-V - 1905105PR; 100202H2HS-V - 1903155UR, 1904105UR, 1903185WR, 1903205WR, 1904175WR, 1905065WR, 1905135WR, and 1905155WR; 100203H2HS-V - 1903015UR, 1904015ZR, 1904105UR, 1904155UR, 1904175UR, 1904225UR, 1905025UR, 1905035NR, 1905035UR, 1905065UR, 1905085UR, 1905135NR, 1905155NR, 1903185WR, and 1905035WR; 100204H2HS-V - 1903015UR, 1903065UR, 1903085UR, 1904035UR, 1904155UR, 1904175UR, 1904175ZR, 1904225UR, 1904245UR, 1904245ZR, 1904265UR, 1904295UR, 1905025UR, 1905035NR, 1905035UR, 1905065UR, 1905085UR, 1905135NR, 1905155NR, 1904105WR, 1904175WR, 1905035WR, 1905065WR, 1905085WR, 1905135WR, and 1905155WR; 100206H2HS-V - 1903045UR, 1903065UR, 1903205UR, 1903225UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR, 1904155UR, 1904175UR, 1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR, 1905035UR, 1905065NR, 1905065UR, 1905085UR, 1905105UR, 1905135UR, 1903185WR, 1904225WR, 1905065WR, 1905105PR, and 1905135WR; 100208H2HS-V - 1903015UR, 1903045UR, 1903085UR, 1903205UR, 1903225UR, 1904015UR 1904035UR, 1904175ZR, 1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR, 1905035UR, 1905065UR, 1905085UR, 1905105UR, 1905135UR, and 1905085WR; 100254H2HS-V - 1904245NR, 1903185WR, and 1905065WR; 100256H2HS-V - 1905065WR; 100304H2HS-V - 1903065WR, 1903205WR, and 1904225WR; 100306H2HS-V - 1903065UR, 1903085UR, 1904155UR, 1904175UR, 1904225UR, 1905135NR, 1905035WR, 1905065WR, 1905105PR, 1905135WR, and 1905155WR; 100308H2HS-V - 1904015UR, 1904155UR, 1904175UR, 1904225UR, 1904265UR, 1904295UR, 1905025UR, 1905065NR, 1905085UR, 1905105UR, 1905135NR, 1905155NR, 1903065WR, 1903185WR, 1903205WR, 1905065WR, 1905105WR, 1905155WR, and 1905225WR; 100310H2HS-V - 1903135UR, 1903155UR, 1903185UR, 1903205UR, 1904015UR, 1904035UR, 1904055UR, 1904085UR, 1904155UR, 1904175UR, 1904175ZR, 1904225UR, 1904245UR, 1904295UR, 1905085WR, and 1905225WR; 100404H2HS-V - 1904175WR, 1904225WR, 1905035WR, 1905065WR, and 1905155WR; 100406H2HS-V - 1903205WR, 1904175WR, and 1905035WR; 100408H2HS-V - 1903085UR, 1903135UR, 1903155UR, 1903155ZR, 1903185UR, 1903225UR,1904015UR, 1904035UR, 1904055UR, 1904225UR, 1903155PR, 1903205WR, 1904105WR, 1904245WR, and 1905065WR; 100410H2HS-V - 1903115UR, 1903155UR, 1903205UR, 1904155UR, 1904175UR, 1904225UR, 1903045WR, and 1905085WR; 100506H2HS-V - 1905155WR; 100508H2HS-V - 1903205WR; 100510H2HS-V - 1903045WR, 1905085WR, and 1905225WR; 100515H2HS-V - 1903065WR, 1903225W6, 1904175WR, 1905085WR, and 1905225WR; 100520H2HS-V - 1904155WR, 1904175WR, 1905085WR, and 1905225WR; 100620H2HS-V - 1903045WR, 1904175WR, 1904225WR, 1905085WR; and MV-00208HHSA - 1904175YR, 1905065XR, and . Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The devices may be missing the implant coil.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Microvention, Inc.
Distributor
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
Lot
Catalog/lot numbers: 100101H2HS-V - 1903135UR, 1903185UR, 1903225UR, and 1904035UR; 100102H2HS - V -1903135UR, 1903155UR, 1903155ZR, 1903185UR, 1903205UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR,1904155UR, 1904175UR, 1904175ZR, and 1904245UR; 100103H2HS-V - 1903015UR, 1903045UR, 1903065UR, 1903085UR, 1903135UR, 1903205UR 1904055UR, 1904085UR, 1904155UR, 1905085UR, 1905105UR, 1905135UR, and 1905155UR; 100104H2HS-V - 1903015UR, 1903135UR, 1903205UR, 1904085UR, 1904105UR, and 1904155UR; 100151H2HS-V - 1903085UR, 1903185UR, and 1904015UR; 100152H2HS-V - 1903135UR, 1903155UR, 1903205NR, 1903205UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR, 1904155UR, 1904175UR, 1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR, 1905035UR, 1905065UR, 1905085UR, 1903045WR, 1903185WR, 1903205WR, 1904105WR, 1904175WR, 1904225WR, 1905065WR, and 1905175WR; 100153H2HS-V - 1903015UR, 1903065UR, 1903085UR, 1903185UR, 1904015UR, 1904015ZR, 1904105UR, 1904155UR, 1904245UR, 1904265UR, and 1904295UR; 100154H2HS-V - 1903135UR, 1903155UR, 1903205UR, 1903225UR, 1904015UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR, 1904155UR, 1904175UR, 1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR, 1905035UR, 1905065NR, 1905065UR, 1905155NR, 1903185WR, 1903205WR, 1904105WR, 1904225WR, and 1905065WR; 100201H2HS-V - 1905105PR; 100202H2HS-V - 1903155UR, 1904105UR, 1903185WR, 1903205WR, 1904175WR, 1905065WR, 1905135WR, and 1905155WR; 100203H2HS-V - 1903015UR, 1904015ZR, 1904105UR, 1904155UR, 1904175UR, 1904225UR, 1905025UR, 1905035NR, 1905035UR, 1905065UR, 1905085UR, 1905135NR, 1905155NR, 1903185WR, and 1905035WR; 100204H2HS-V - 1903015UR, 1903065UR, 1903085UR, 1904035UR, 1904155UR, 1904175UR, 1904175ZR, 1904225UR, 1904245UR, 1904245ZR, 1904265UR, 1904295UR, 1905025UR, 1905035NR, 1905035UR, 1905065UR, 1905085UR, 1905135NR, 1905155NR, 1904105WR, 1904175WR, 1905035WR, 1905065WR, 1905085WR, 1905135WR, and 1905155WR; 100206H2HS-V - 1903045UR, 1903065UR, 1903205UR, 1903225UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR, 1904155UR, 1904175UR, 1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR, 1905035UR, 1905065NR, 1905065UR, 1905085UR, 1905105UR, 1905135UR, 1903185WR, 1904225WR, 1905065WR, 1905105PR, and 1905135WR; 100208H2HS-V - 1903015UR, 1903045UR, 1903085UR, 1903205UR, 1903225UR, 1904015UR 1904035UR, 1904175ZR, 1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR, 1905035UR, 1905065UR, 1905085UR, 1905105UR, 1905135UR, and 1905085WR; 100254H2HS-V - 1904245NR, 1903185WR, and 1905065WR; 100256H2HS-V - 1905065WR; 100304H2HS-V - 1903065WR, 1903205WR, and 1904225WR; 100306H2HS-V - 1903065UR, 1903085UR, 1904155UR, 1904175UR, 1904225UR, 1905135NR, 1905035WR, 1905065WR, 1905105PR, 1905135WR, and 1905155WR; 100308H2HS-V - 1904015UR, 1904155UR, 1904175UR, 1904225UR, 1904265UR, 1904295UR, 1905025UR, 1905065NR, 1905085UR, 1905105UR, 1905135NR, 1905155NR, 1903065WR, 1903185WR, 1903205WR, 1905065WR, 1905105WR, 1905155WR, and 1905225WR; 100310H2HS-V - 1903135UR, 1903155UR, 1903185UR, 1903205UR, 1904015UR, 1904035UR, 1904055UR, 1904085UR, 1904155UR, 1904175UR, 1904175ZR, 1904225UR, 1904245UR, 1904295UR, 1905085WR, and 1905225WR; 100404H2HS-V - 1904175WR, 1904225WR, 1905035WR, 1905065WR, and 1905155WR; 100406H2HS-V - 1903205WR, 1904175WR, and 1905035WR; 100408H2HS-V - 1903085UR, 1903135UR, 1903155UR, 1903155ZR, 1903185UR, 1903225UR,1904015UR, 1904035UR, 1904055UR, 1904225UR, 1903155PR, 1903205WR, 1904105WR, 1904245WR, and 1905065WR; 100410H2HS-V - 1903115UR, 1903155UR, 1903205UR, 1904155UR, 1904175UR, 1904225UR, 1903045WR, and 1905085WR; 100506H2HS-V - 1905155WR; 100508H2HS-V - 1903205WR; 100510H2HS-V - 1903045WR, 1905085WR, and 1905225WR; 100515H2HS-V - 1903065WR, 1903225W6, 1904175WR, 1905085WR, and 1905225WR; 100520H2HS-V - 1904155WR, 1904175WR, 1905085WR, and 1905225WR; 100620H2HS-V - 1903045WR, 1904175WR, 1904225WR, 1905085WR; and MV-00208HHSA - 1904175YR, 1905065XR, and
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was MicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, Helical, Sterile, Rx, REF numbers 100101H2HS-V; 100102H2HS-V, 100103H2HS-V, 100104H2HS-V, 100151H2HS-V; 100152H2HS-V; 100153H2HS-V; 100154H2HS-V; 100201H2HS-V; 100202H2HS-V; 100203H2HS-V; 100204H2HS-V; 100206H2HS-V, 100208H2HS-V, 100254H2HS-V, 100256H2HS-V, 100304H2HS-V; 100306H2HS-V; 100308H2HS-V; 100310H2HS-V; 100404H2HS-V; 100406H2HS-V, 100408H2HS-V, 100410H2HS-V, 100506H2HS-V, 100508H2HS-V; 100510H2HS-V, 100515H2HS-V, 100520H2HS-V, 100620H2HS-V, and MV-00208HHSA - Product Usage: The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA. recalled?

Yes. FDA published recall Z-1392-2020 on March 4, 2020. Reported hazard: The devices may be missing the implant coil.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail