Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-502 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to the Operating Room, in dental clinics or hospital wards to clean, thermally or chemically disinfect and dry surgical goods.
Published March 4, 2020 · Recall initiated: January 16, 2020
Summary
FDA records medical device Class III recall Z-1390-2020, published March 4, 2020 and initiated January 16, 2020. Recalling firm: Maquet Cardiovascular Us Sales, Llc; Status: Terminated; Reported quantity: 4 units; Lot: Serial Numbers: WAA090630, WAA090724, WAA091212, WAA091456, WAA091468, WAA092596. UDI Code 07340153700031. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a missing water intake for Deionized water on the Getinge 46-Series Washer Disinfector. There were no valve, hose, or clamps for this option included with the device therefore the units could not be properly installed. There have been no adverse events reported resulting in serious illness or injuries.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Cardiovascular Us Sales, Llc
- Distributor
- U.S. Distribution: CA, CO, IA, MN, NY, SC, VA
- Lot
- Serial Numbers: WAA090630, WAA090724, WAA091212, WAA091456, WAA091468, WAA092596. UDI Code 07340153700031
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHLS Set Advanced, Model: BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Product Code/Part Number: 70105.2794 - ProductDecember 23, 2020
- FDAHLS Set Advanced Model Name: BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code/Part Number: 70105.2797 -December 23, 2020
- FDAYuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before,October 28, 2020
- FDAMaquet/Getinge Rotaflow System Console US (115 V / 60 Hz) - Product Usage: Instructions for Use - CentrifugalSeptember 23, 2020
- FDACERTAIN BELLATEK ENCODE HEALING ABUTMENT 5MM(D) X 6MM(P) X 6MM(H) Reference Number: IEHA566March 4, 2020
- FDAMicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, Helical, Sterile,March 4, 2020
Questions about this recall
Was Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-502 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to the Operating Room, in dental clinics or hospital wards to clean, thermally or chemically disinfect and dry surgical goods. recalled?
Yes. FDA published recall Z-1390-2020 on March 4, 2020. Reported hazard: Due to a missing water intake for Deionized water on the Getinge 46-Series Washer Disinfector. There were no valve, hose, or clamps for this option included with the device therefore the units could not be properly installed. There have been no adverse events reported resulting in serious illness or injuries.
What should I do if I have this product?
Voluntary: Firm initiated · Letter