Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, 2.1mm Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for Leksell Stereotactic System. Each inner box of REF 911761 contains a Tyvek bag labeled REF 912465, Disposable Biopsy Needle and Aspirator Tube.
Published April 24, 2024 · Recall initiated: March 15, 2024
Summary
FDA records medical device Class I recall Z-1397-2024, published April 24, 2024 and initiated March 15, 2024. Recalling firm: Elekta Instrument AB; Status: Ongoing; Reported quantity: 601 needles; Lot: Lot number 837838839 appears on the labels for REF 911933, REF 911761, and REF 912465. UDI numbers: REF 911933 - (01)0 7340048 30034 6, (17)250401 (10) 837838839 (240) 911933; REF 911761 - (01)0 7340048 30800 7, (17)250401 (10) 837838839 (240) 911761; REF 912465 - (01)0 7340048 30801 4, (17)250401 (10) 837838839 (240) 912465. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Elekta has become aware that Disposable Biopsy Needles (911933) from one batch (837838839) can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles. The sterility of the biopsy needles has not been affected. This issue has been reported from one site.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Elekta Instrument AB
- Distributor
- US and Australia, Austria, Belgium, France, Germany, Greece, Italy, Netherlands, Spain, Sweden, Switzerland
- Lot
- Lot number 837838839 appears on the labels for REF 911933, REF 911761, and REF 912465. UDI numbers: REF 911933 - (01)0 7340048 30034 6, (17)250401 (10) 837838839 (240) 911933; REF 911761 - (01)0 7340048 30800 7, (17)250401 (10) 837838839 (240) 911761; REF 912465 - (01)0 7340048 30801 4, (17)250401 (10) 837838839 (240) 912465.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- Sweden
Related recalls
- FDAElekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge,November 13, 2024
- FDAHelados Mexico Premium Ice Cream Minis Variety Pack 12 bars Strawberry, Coconut, Mango 1.5 FL OZ (44mL) BardsApril 24, 2024
- FDATablo Hemodialysis System, REF: PN-0003000, PN-0006000 and PN-0007001April 24, 2024
- FDATranquility & Teachings William Monague plate set, brand Indigenous Collection by CAP. Each box consists ofApril 24, 2024
- FDAAeris EVO, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outsideApril 24, 2024
- FDALeksell GammaPlan 11.1 Product Usage: Leksell GammaPlan is a computer-based system designed for Leksell GammaMay 29, 2019
Questions about this recall
Was Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, 2.1mm Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for Leksell Stereotactic System. Each inner box of REF 911761 contains a Tyvek bag labeled REF 912465, Disposable Biopsy Needle and Aspirator Tube. recalled?
Yes. FDA published recall Z-1397-2024 on April 24, 2024. Reported hazard: Elekta has become aware that Disposable Biopsy Needles (911933) from one batch (837838839) can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles. The sterility of the biopsy needles has not been affected. This issue has been reported from one site.
What should I do if I have this product?
Voluntary: Firm initiated