AURIGA 30 GENERAL ZERO COST SYSTEM UPN: M0068S30GZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.
Published May 29, 2019 · Recall initiated: April 8, 2019
Summary
FDA records medical device Class II recall Z-1399-2019, published May 29, 2019 and initiated April 8, 2019. Recalling firm: Boston Scientific Corporation; Status: Terminated; Reported quantity: 2 (OUS); Lot: Serial Number: 21302836, 21302929. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Boston Scientific Corporation
- Distributor
- Worldwide distribution to Australia, Belgium, Brazil, Chile, Colombia, Egypt, France, Germany, Great Britain, India, Indonesia, Iran, Italy, Lebanon, Malaysia, Mexico, Romania, Russia Fed., South Africa, South Korea, Spain, Taiwan, United Arab Emirates and Vietnam
- Lot
- Serial Number: 21302836, 21302929
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;October 21, 2026
- FDAImager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/038 BX5;September 30, 2026
- FDABoston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVISeptember 23, 2026
- FDALUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model NumberSeptember 9, 2026
- FDAAURIGA 30 BRAZIL ZERO COST SYSTEM UPN: M0068S30BZ0 Product Usage: The Auriga XL 4007 Laser System and theMay 29, 2019
- FDALeksell GammaPlan 11.1 Product Usage: Leksell GammaPlan is a computer-based system designed for Leksell GammaMay 29, 2019
Questions about this recall
Was AURIGA 30 GENERAL ZERO COST SYSTEM UPN: M0068S30GZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery. recalled?
Yes. FDA published recall Z-1399-2019 on May 29, 2019. Reported hazard: Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.
What should I do if I have this product?
Voluntary: Firm initiated · Letter