MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft Helical, Sterile, Rx, REF numbers 100101HS-V, 100102HS-V, 100103HS-V, 100104HS-V, 100151HS-V, 100152HS-V, 100153HS-V, 100154HS-V, 100201HS-V, 100202HS-V, 100203HS-V, 100204HS-V, 100206HS-V, 100254HS-V, 100256HS-V, 100304HS-V, 100306HS-V, 100308HS-V, 100406HS-V, 100408HS-V, and 100508HS-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.

Published March 4, 2020 · Recall initiated: November 22, 2019

Summary

FDA records medical device Class II recall Z-1407-2020, published March 4, 2020 and initiated November 22, 2019. Recalling firm: Microvention, Inc.; Status: Terminated; Reported quantity: 5,777 units; Lot: Catalog/lot numbers: 100101HS-V - 1903045UH, 1903065UH, 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1903225ZH, 1904085UH, 1904105UH, 1904175UH, 1904245UH, 1905025UH, 1905085UH, 1905105ZH, 1905135UH, 1905175UH, 1905205UH, 1905205ZH, 1905225UH, 1905225ZH, 1903135WH, 1904175WH, 1905085WH, 1905105PH, and 1905135WH; 100102HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1903225ZH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1905205ZH, 1905225UH, 1905225ZH, 1904035WH, and 1905155WH; 100103HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1904105UH, 1904155UH, 1904175UH, 1904225UH, 1904245UH, 1905025UH, 1905025ZH, 1905035UH, 1905035ZH, 1905065UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, and 1905085WH; 100104HS-V - 1903115UH, 1903135UH, 1903225UH, 1904085UH, 1904105UH, 1904155UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1904175WH, 1904225WH, and 1905155WH; 100151HS-V - 1905025UH, 1903135TH, 1903085WH, 1903205WH, 1904015WH, 1904265WH, 1904295WH, 1905085WH, 1905155WH, and 1905225PH; 100152HS-V - 1903155UH, 1903185UH, 1903205UH, 1904085UH, 1904105UH, 1904155UH, 1904225UH, 1904245UH, 1905025UH, 1905025ZH, 1905035UH, 1905035ZH, 1905065UH. 1905085UH, 1905105UH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205ZH, 1903065NH, 1904175TH, 1904225NH, 1905065NH, 1903015WH, 1903065WH, 1903185WH, 1904015WH, 1904035WH, 1904055WH, 1904085WH, 1904155WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035PH, 1905035WH, 1905065WH, 1905085WH, 1905155WH, and 1905225WH; 100153HS-V - 1903225UH, 1905105UH, 1905135UH, 1905135ZH, 1905155UH, 1905175UH, 1905205UH, 1905225UH, 1905225ZH, 1903015NH, 1903065WH, 1903085WH, 1903205WH, 1904015WH, 1904035WH, 1904055WH, 1904155WH, 1904175WH, 1904225PH, 1904225WH, 1904245WH, 1904265WH, 1905025WH, 1905035PH, 1905065WH, 1905085WH, 1905155WH, 1905205PH, 1905225PH, and 1905225WH. 100154HS-V - 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1905225UH, 1903115TH, 1903015WH, 1903045WH, 1903065WH, 1904035WH, 1904055WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905035PH, 1905035WH, 1905065WH, 1905085WH, 1905155PH, 1905155WH, 1905175WH, and 1905205WH; 100201HS-V - 1903135UH 100202HS-V - 1904085UH, 1904105UH, 1904175UH, 1904245UH, 1904225NH, 1905065NH, 1905205NH, 1903085WH, 1904015WH, 1904055WH, 1904175WH, 1904245WH, 1904265WH, 1904295WH, 1905085WH, 1905225PH, and 1905225WH; 100203HS-V - 1903155UH, 1903185UH, 1903225UH, 1904225NH, 1905135NH, 1903015WH, 1903065WH, 1904035WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905035WH, 1905155WH, and 1905225WH; 100204HS-V - 1904085UH, 1904105UH, 1904225UH, 1904245UH, 1905065UH, 1905085UH, 1905105UH, 1905135UH, 1905135ZH, 1905155UH, 1905175UH, 1905205UH, 1903045WH, 1903065WH, 1904015WH, 1904035WH, 1904155WH, 1904175WH, 1904225PH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035WH, 1905085WH, 1905155WH, and 1905225WH; 100206HS-V - 1904085UH, 1904105UH, 1904155UH, 1904175UH, 1904265UH, 1905065UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905175UH, 1905205UH 1905205ZH, 1905225UH, 1903065WH, 1903085WH, 1903185WH, 1904055WH, 1904175WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035PH, 1905035WH, 1905065WH, 1905155WH, 1905175WH, and 1905225WH; 100254HS-V - 1903205WH, 1904015WH, 1904225WH, 1904265WH, 1904295WH, 1905025WH, 1905085WH, and 1905225PH; 100256HS-V - 1903065WH, 1903185WH, 1904015WH, and 1905155WH; 100304HS-V - 1903085NH, 1903135TH, 1904035TH, 1903065WH, 1905035PH, and 1905065WH; 100306HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1904085UH, 1903065WH, 1904035WH, 1904055WH, 1904155WH, 1904175WH, 1904225WH, 1904245WH, 1905035PH, 1905065WH, 1905205PH, and 1905225PH; 100308HS-V - 1903185UH, 1903225UH. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The devices may be missing the implant coil.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Microvention, Inc.
Distributor
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
Lot
Catalog/lot numbers: 100101HS-V - 1903045UH, 1903065UH, 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1903225ZH, 1904085UH, 1904105UH, 1904175UH, 1904245UH, 1905025UH, 1905085UH, 1905105ZH, 1905135UH, 1905175UH, 1905205UH, 1905205ZH, 1905225UH, 1905225ZH, 1903135WH, 1904175WH, 1905085WH, 1905105PH, and 1905135WH; 100102HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1903225ZH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1905205ZH, 1905225UH, 1905225ZH, 1904035WH, and 1905155WH; 100103HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1904105UH, 1904155UH, 1904175UH, 1904225UH, 1904245UH, 1905025UH, 1905025ZH, 1905035UH, 1905035ZH, 1905065UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, and 1905085WH; 100104HS-V - 1903115UH, 1903135UH, 1903225UH, 1904085UH, 1904105UH, 1904155UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1904175WH, 1904225WH, and 1905155WH; 100151HS-V - 1905025UH, 1903135TH, 1903085WH, 1903205WH, 1904015WH, 1904265WH, 1904295WH, 1905085WH, 1905155WH, and 1905225PH; 100152HS-V - 1903155UH, 1903185UH, 1903205UH, 1904085UH, 1904105UH, 1904155UH, 1904225UH, 1904245UH, 1905025UH, 1905025ZH, 1905035UH, 1905035ZH, 1905065UH. 1905085UH, 1905105UH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205ZH, 1903065NH, 1904175TH, 1904225NH, 1905065NH, 1903015WH, 1903065WH, 1903185WH, 1904015WH, 1904035WH, 1904055WH, 1904085WH, 1904155WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035PH, 1905035WH, 1905065WH, 1905085WH, 1905155WH, and 1905225WH; 100153HS-V - 1903225UH, 1905105UH, 1905135UH, 1905135ZH, 1905155UH, 1905175UH, 1905205UH, 1905225UH, 1905225ZH, 1903015NH, 1903065WH, 1903085WH, 1903205WH, 1904015WH, 1904035WH, 1904055WH, 1904155WH, 1904175WH, 1904225PH, 1904225WH, 1904245WH, 1904265WH, 1905025WH, 1905035PH, 1905065WH, 1905085WH, 1905155WH, 1905205PH, 1905225PH, and 1905225WH. 100154HS-V - 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1905225UH, 1903115TH, 1903015WH, 1903045WH, 1903065WH, 1904035WH, 1904055WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905035PH, 1905035WH, 1905065WH, 1905085WH, 1905155PH, 1905155WH, 1905175WH, and 1905205WH; 100201HS-V - 1903135UH 100202HS-V - 1904085UH, 1904105UH, 1904175UH, 1904245UH, 1904225NH, 1905065NH, 1905205NH, 1903085WH, 1904015WH, 1904055WH, 1904175WH, 1904245WH, 1904265WH, 1904295WH, 1905085WH, 1905225PH, and 1905225WH; 100203HS-V - 1903155UH, 1903185UH, 1903225UH, 1904225NH, 1905135NH, 1903015WH, 1903065WH, 1904035WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905035WH, 1905155WH, and 1905225WH; 100204HS-V - 1904085UH, 1904105UH, 1904225UH, 1904245UH, 1905065UH, 1905085UH, 1905105UH, 1905135UH, 1905135ZH, 1905155UH, 1905175UH, 1905205UH, 1903045WH, 1903065WH, 1904015WH, 1904035WH, 1904155WH, 1904175WH, 1904225PH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035WH, 1905085WH, 1905155WH, and 1905225WH; 100206HS-V - 1904085UH, 1904105UH, 1904155UH, 1904175UH, 1904265UH, 1905065UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905175UH, 1905205UH 1905205ZH, 1905225UH, 1903065WH, 1903085WH, 1903185WH, 1904055WH, 1904175WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035PH, 1905035WH, 1905065WH, 1905155WH, 1905175WH, and 1905225WH; 100254HS-V - 1903205WH, 1904015WH, 1904225WH, 1904265WH, 1904295WH, 1905025WH, 1905085WH, and 1905225PH; 100256HS-V - 1903065WH, 1903185WH, 1904015WH, and 1905155WH; 100304HS-V - 1903085NH, 1903135TH, 1904035TH, 1903065WH, 1905035PH, and 1905065WH; 100306HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1904085UH, 1903065WH, 1904035WH, 1904055WH, 1904155WH, 1904175WH, 1904225WH, 1904245WH, 1905035PH, 1905065WH, 1905205PH, and 1905225PH; 100308HS-V - 1903185UH, 1903225UH
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft Helical, Sterile, Rx, REF numbers 100101HS-V, 100102HS-V, 100103HS-V, 100104HS-V, 100151HS-V, 100152HS-V, 100153HS-V, 100154HS-V, 100201HS-V, 100202HS-V, 100203HS-V, 100204HS-V, 100206HS-V, 100254HS-V, 100256HS-V, 100304HS-V, 100306HS-V, 100308HS-V, 100406HS-V, 100408HS-V, and 100508HS-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA. recalled?

Yes. FDA published recall Z-1407-2020 on March 4, 2020. Reported hazard: The devices may be missing the implant coil.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail