HeartStart MRx Monitor/Defibrillator, Model numbers: M3535A (M3535ATZ) M3536A (M3536ATZ), M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, M3536M9
Published April 5, 2017 · Recall initiated: February 24, 2017
Summary
FDA records medical device Class I recall Z-1417-2017, published April 5, 2017 and initiated February 24, 2017. Recalling firm: Philips North America, LLC; Status: Terminated; Reported quantity: 97,794 total ( 47,362 US and 50,429 OUS); Lot: Model# Serial Numbers M3535A US00100100 to US00602546 M3535ATZ US00100100 to US00602546 M3536A US00100902 to US00602540 (M3536ATZ) US00100902 to US00602540 M3536M US00500002 to US00553553 M3536MC US00500001 to US00500087 M3536M2 US00554176 to US00554178 M3536M3 US00571854 to US00588995 M3536M4 US00500003 to US00590982 M3536M5 US00500001 to US00587771 M3536M6 US00554358 to US00600582 M3536M7 US00581047 to US00601962 M3536M8 US00590958 to US0059095 M3536M9 US00581048 to US00581049. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
HeartStart MRx Monitor/Defibrillator may may exhibit abnormal behavior when operating using a single battery as the sole source of power may fail to deliver therapy or monitoring
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America, LLC
- Distributor
- Worldwide Distribution-US (Nationwide) including PUERTO RICO and countries of: AFGHANISTAN, ALBANIA, ANGOLA, ARGENTINA, ARUBA, AUSTRALIA,AUSTRIA,AZERBAIJAN,BAHRAIN, BANGLADESH, BELGIUM, BOLIVIA,BOSNIA AND HERZEGOWINA, BRAZIL,BRUNEI DARUSSALAM, BULGARIA,CAMBODIA,CANADA,CHILE,CHINA,COLOMBIA,COSTA RICA, CROATIA (local name: Hrvatska),CYPRUS,CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC,ECUADOR,EGYPT,EL SALVADOR, ESTONIA, ETHIOPIA,FINLAND,FRANCE,GABON,GERMANY,GHANA,GREECE, GUADELOUPE, HONG KONG,HUNGARY,ICELAND,INDIA, INDONESIA,IRAN (ISLAMIC REPUBLIC OF),IRAQ,IRELAND,ISRAEL,ITALY,JAPAN, JORDAN, KAZAKHSTAN, KENYA,KOREA, REPUBLIC OF,KUWAIT,LAOS,LATVIA, LEBANON, LIBYAN ARAB JAMAHIRIYA,LITHUANIA,LUXEMBOURG, MACAU, MACEDONIA,MALAYSIA,MALTA,MAURITIUS,MEXICO,MONGOLIA, MOROCCO, MOZAMBIQUE,MYANMAR (Burma), NAMIBIA, NEPAL, NETHERLANDS, NETHERLANDS ANTILLES,NEW CALEDONIA,NEW ZEALAND, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PANAMA,PAPUA NEW GUINEA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL,QATAR,REUNION,ROMANIA,RUSSIAN FEDERATION,SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA,SOUTH AFRICA,SPAIN,SRI LANKA, SWEDEN,SWITZERLAND,TAIWAN, PROVINCE OF CHINA,TANZANIA, UNITED REPUBLIC OF,THAILAND,TRINIDAD AND TOBAGO, TUNISIA, TURKEY, TURKMENISTAN,UGANDA,UKRAINE,UNITED ARAB EMIRATES,UNITED KINGDOM,UZBEKISTAN,VENEZUELA,VIET NAM,YEMEN.
- Lot
- Model# Serial Numbers M3535A US00100100 to US00602546 M3535ATZ US00100100 to US00602546 M3536A US00100902 to US00602540 (M3536ATZ) US00100902 to US00602540 M3536M US00500002 to US00553553 M3536MC US00500001 to US00500087 M3536M2 US00554176 to US00554178 M3536M3 US00571854 to US00588995 M3536M4 US00500003 to US00590982 M3536M5 US00500001 to US00587771 M3536M6 US00554358 to US00600582 M3536M7 US00581047 to US00601962 M3536M8 US00590958 to US0059095 M3536M9 US00581048 to US00581049
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe IFU for the Autoclavable Temperature Probes: Part # 453564635891, providing in the packaging for theApril 20, 2022
- FDAPhilips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A,November 4, 2020
- FDAPhilips SureSigns VM1SpO2, C02, Rec, model no. 863266 - Product Usage: The SureSigns VM1 is a patient monitorJune 17, 2020
- FDAPhilips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospitalMay 27, 2020
- FDABaidyanath brand Chitrakadi Bati, Dietary Supplement, 80 Tablets -- Product of India --- Mfd in India - ShreeApril 5, 2017
- FDAXTRAC Excimer Laser, Model No. AL8000, AL10000 The intended use is targeted UVB phototherapy for treatment ofApril 5, 2017
Questions about this recall
Was HeartStart MRx Monitor/Defibrillator, Model numbers: M3535A (M3535ATZ) M3536A (M3536ATZ), M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, M3536M9 recalled?
Yes. FDA published recall Z-1417-2017 on April 5, 2017. Reported hazard: HeartStart MRx Monitor/Defibrillator may may exhibit abnormal behavior when operating using a single battery as the sole source of power may fail to deliver therapy or monitoring
What should I do if I have this product?
Voluntary: Firm initiated · Letter