ADVIA Centaur Progesterone; Catalog Number: 10310305 10315522 10333111 Product Usage: The PROG method is an in vitro diagnostic test for the quantitative measurement of progesterone in human serum and plasma on the Dimension Vista®System. Progesterone measurements are used in the diagnosis and treatment of disorders of the ovaries or placenta.
Published March 15, 2017 · Recall initiated: January 4, 2017
Summary
FDA records medical device Class II recall Z-1418-2017, published March 15, 2017 and initiated January 4, 2017. Recalling firm: Siemens Healthcare Diagnostics, Inc; Status: Terminated; Reported quantity: 76,107; Lot: SMN: 10310305: 75422274 83380274 89389274 99473274 8713275 19796275 30565275 38674275 46515276 55704276 66971276 78934276 80930277; SMN: 10315522: 74355274 80476274 87497274 20274 7772275 14822275 25943275 35303275 43229276 46525276 55505276 67430276 80448276 81415277; SMN: 10333111: 84848274 89946274 14006275 33598275 49907276 66978276. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
DHEA-S causes falsely elevated progesterone results.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc
- Distributor
- US Nationwide distribution including Puerto Rico
- Lot
- SMN: 10310305: 75422274 83380274 89389274 99473274 8713275 19796275 30565275 38674275 46515276 55704276 66971276 78934276 80930277; SMN: 10315522: 74355274 80476274 87497274 20274 7772275 14822275 25943275 35303275 43229276 46525276 55505276 67430276 80448276 81415277; SMN: 10333111: 84848274 89946274 14006275 33598275 49907276 66978276
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASiemens ADVIA Centaur Folate 100 test kit-for IVD of folate in serum or red blood cells SMN:10310308January 29, 2020
- FDASiemensADVIA Centaur aTG (500 test kit) Siemens Material Number (SMN): 10492399May 8, 2019
- FDAAtellica IM aTG (500 test kit) Siemens Material Number (SMN): 10995462May 8, 2019
- FDASiemens Atellica IM aTG (100 test kit) Siemens Material Number (SMN): 10995461May 8, 2019
- FDAEpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with theMarch 15, 2017
- FDADimension Vista LOCI Progesterone; Catalog Number: K6464; Siemens Material Number: 10461743 Product Usage: ForMarch 15, 2017
Questions about this recall
Was ADVIA Centaur Progesterone; Catalog Number: 10310305 10315522 10333111 Product Usage: The PROG method is an in vitro diagnostic test for the quantitative measurement of progesterone in human serum and plasma on the Dimension Vista®System. Progesterone measurements are used in the diagnosis and treatment of disorders of the ovaries or placenta. recalled?
Yes. FDA published recall Z-1418-2017 on March 15, 2017. Reported hazard: DHEA-S causes falsely elevated progesterone results.
What should I do if I have this product?
Voluntary: Firm initiated · Letter