Juno DRF; Model: 709020 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures
Published March 15, 2017 · Recall initiated: December 27, 2016
Summary
FDA records medical device Class II recall Z-1427-2017, published March 15, 2017 and initiated December 27, 2016. Recalling firm: Villa Sistemi Medicali S.P.A.; Status: Terminated; Reported quantity: 65 devices in total; Lot: Serial Numbers: 10110770 11030805 11070845 11070848 11070849 11090864 11110900 12020978 12020982 12030999 12041003 12061014 12061018 12061020 12081040 12091074 12101104 12121186 12121187 13031215 13101263 14021301 14031313 14041321 14051332 14071353 14081359 14091368 14091371 14121400 14121403 15021426 15031440 15041444 15091473 15091478 15101484 15101487 15121505 16011512 16021514 16021517 16021518 16041533 16121595. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Villa Sistemi Medicali S.P.A.
- Distributor
- US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, PA, Puerto Rico, TN, TX, UT, and W.I
- Lot
- Serial Numbers: 10110770 11030805 11070845 11070848 11070849 11090864 11110900 12020978 12020982 12030999 12041003 12061014 12061018 12061020 12081040 12091074 12101104 12121186 12121187 13031215 13101263 14021301 14031313 14041321 14051332 14071353 14081359 14091368 14091371 14121400 14121403 15021426 15031440 15041444 15091473 15091478 15101484 15101487 15121505 16011512 16021514 16021517 16021518 16041533 16121595
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Italy
Related recalls
- FDAJuno DRF, Model Number 709020 - Product Usage: General. multi-function x-ray systems capable of routineJanuary 8, 2020
- FDAApollo; Models: 9784000131 9784000231 9784020131 9784020231 9784100231 9784120131 9784120231 9784130131March 15, 2017
- FDAApollo DRF; Model: 9784220831 Product Usage: General, multi-function x-ray systems capable of routineMarch 15, 2017
- FDABiograph 16 TruePoint - 3R, Material Number 10249555 The Biograph Mobile PET/CTs are intended to be utilizedMarch 15, 2017
- FDABiograph 6 TruePoint, Material Number 10097289 The Biograph Mobile PET/CTs are intended to be utilized byMarch 15, 2017
- FDAIMMULITE 2000 Progesterone; Catalog Number: L2KPW2 L2KPW6; Siemens Material Number: 10381181, 10381170 ProductMarch 15, 2017
Questions about this recall
Was Juno DRF; Model: 709020 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures recalled?
Yes. FDA published recall Z-1427-2017 on March 15, 2017. Reported hazard: Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.
What should I do if I have this product?
Voluntary: Firm initiated · Letter