8mm Maryland Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.
Published June 5, 2013 · Recall initiated: May 8, 2013
Summary
FDA records medical device Class II recall Z-1432-2013, published June 5, 2013 and initiated May 8, 2013. Recalling firm: Intuitive Surgical, Inc.; Status: Terminated; Reported quantity: 697 - all products; Lot: Model number: 400172-08; Lot numbers: M10120815 M10120817 M10120831 M10120913 M10120918 M10121008 M10121024 M10121120 M10121123 M10121203 M10121210 M10121215 M10130220 M10130222 M10130311 M10130315 M10130327 M10130411. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Intuitive Surgical, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Saudi Arabia, Singapore, and Venezuela.
- Lot
- Model number: 400172-08; Lot numbers: M10120815 M10120817 M10120831 M10120913 M10120918 M10121008 M10121024 M10121120 M10121123 M10121203 M10121210 M10121215 M10130220 M10130222 M10130311 M10130315 M10130327 M10130411.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15July 22, 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 380601July 1, 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/AJune 10, 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06May 13, 2026
- FDA8mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple useJune 5, 2013
- FDA8mm Fenestrated Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. MultipleJune 5, 2013
Questions about this recall
Was 8mm Maryland Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments. recalled?
Yes. FDA published recall Z-1432-2013 on June 5, 2013. Reported hazard: Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.
What should I do if I have this product?
Voluntary: Firm initiated