8 mm Monopolar Curved Scissors (a.k.a. Hot Shears) Manufactured and Distributed by: Intuitive Surgical Sunny vale, CA The EndoWrist Monopolar Curved Scissors Instrument is a multiple-use endoscopic instrument utilizing a single use, tip cover accessory, to be used in conjunction with the Intuitive Surgical Endoscopic Instrument Control System.
Published June 12, 2013 · Recall initiated: May 8, 2013
Summary
FDA records medical device Class II recall Z-1442-2013, published June 12, 2013 and initiated May 8, 2013. Recalling firm: Intuitive Surgical, Inc.; Status: Terminated; Reported quantity: ~132,441 instruments; Lot: Model Numbers 420179-09 and -10 and 400179-09 and -10. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Intuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and 10 instruments to develop very small cracks near the distal end. This may create a pathway for electrosurgical energy to leak to tissue and potentially cause thermal injury.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Intuitive Surgical, Inc.
- Distributor
- Worldwide distribution - USA (nationwide) Foreign Distribution to Argentina, Austria, Australia, Belgium, Brazil, Bulgaria, Canada, China, Chile, Czech Republic, Denmark, Finland, France, Germany, Cyprus- Greece, Egypt, Guadeloupe, Israel, Ireland, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Norway, Pakistan, Philippines, Portugal, Poland, Qatar, Russia, Romania, Slovenia, South Korea, Spain, Sweden, Switzerland, Thailand, Taiwan, Turkey, United Kingdom, Saudi Arabia, Singapore, South Korea, and Venezuela.
- Lot
- Model Numbers 420179-09 and -10 and 400179-09 and -10.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15July 22, 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 380601July 1, 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/AJune 10, 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06May 13, 2026
- FDAOptovue iVue with Normative Database (NDB) with Software Version 3.0 and 3.1. Intended for in vivo imaging.June 12, 2013
- FDACanister: CO2 canister with GE part numbers 1407-3200-000, 1407-7004-000, or M1084850. Intended to provideJune 12, 2013
Questions about this recall
Was 8 mm Monopolar Curved Scissors (a.k.a. Hot Shears) Manufactured and Distributed by: Intuitive Surgical Sunny vale, CA The EndoWrist Monopolar Curved Scissors Instrument is a multiple-use endoscopic instrument utilizing a single use, tip cover accessory, to be used in conjunction with the Intuitive Surgical Endoscopic Instrument Control System. recalled?
Yes. FDA published recall Z-1442-2013 on June 12, 2013. Reported hazard: Intuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and 10 instruments to develop very small cracks near the distal end. This may create a pathway for electrosurgical energy to leak to tissue and potentially cause thermal injury.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail