Part 510121 Disposable Infusion Pump, ...Elastomeric Pump 300 mL, 4mL/hr,...sterile.. found in the following kits GoPump Kit 510203-BP, Epidural catheter & BIOPATCH GoPump Kit 510204, 2.5" Fenestrated catheter GoPump Kit 510204 BP, 2.5 Fenestrated catheter & Biopatch GoPump Kit 510205BP, 5" Fenestrated catheter & Biopatch GoPump Kit 510551-BP, 2.5 Fenestrated catheter & Biopatch GoBlock kit 510608 Disposable Single Infusion Pump; Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication.

Published June 12, 2013 · Recall initiated: May 10, 2013

Summary

FDA records medical device Class I recall Z-1455-2013, published June 12, 2013 and initiated May 10, 2013. Recalling firm: Symbios Medical Products, LLC; Status: Terminated; Reported quantity: 33,543 in all kits; Lot: Pump Part 510121, Lot 11-100081 to 12-101406 found in: Go Pump Kit 510203-BP. Lot 12-101301 to 13-101538 GoPump Kit 510204, Lot 11-100505 to 11-1-665 GoPump Kit 510204 BP, Lot 11-100721 to 13-100680 GoPump Kit 510205BP, Lot 11-100722 to 13-100087 GoBlock kit 510608. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Symbios Medical Products, LLC
Distributor
US Nationwide distribution.
Lot
Pump Part 510121, Lot 11-100081 to 12-101406 found in: Go Pump Kit 510203-BP. Lot 12-101301 to 13-101538 GoPump Kit 510204, Lot 11-100505 to 11-1-665 GoPump Kit 510204 BP, Lot 11-100721 to 13-100680 GoPump Kit 510205BP, Lot 11-100722 to 13-100087 GoBlock kit 510608
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Part 510121 Disposable Infusion Pump, ...Elastomeric Pump 300 mL, 4mL/hr,...sterile.. found in the following kits GoPump Kit 510203-BP, Epidural catheter & BIOPATCH GoPump Kit 510204, 2.5" Fenestrated catheter GoPump Kit 510204 BP, 2.5 Fenestrated catheter & Biopatch GoPump Kit 510205BP, 5" Fenestrated catheter & Biopatch GoPump Kit 510551-BP, 2.5 Fenestrated catheter & Biopatch GoBlock kit 510608 Disposable Single Infusion Pump; Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication. recalled?

Yes. FDA published recall Z-1455-2013 on June 12, 2013. Reported hazard: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

What should I do if I have this product?

Voluntary: Firm initiated · Letter