Synthes CMF Battery Powered Driver, Part no. 305.840. Used to place screws in the cranialfacial bones, and to drill and/or tap holes to prepare cranialfacial bones for screw insertion.
Published April 23, 2014 · Recall initiated: January 30, 2014
Summary
FDA records medical device Class II recall Z-1456-2014, published April 23, 2014 and initiated January 30, 2014. Recalling firm: Synthes, Inc.; Status: Terminated; Reported quantity: 20081; Lot: Part no. 305.840, with lot nos.: 5680577, 5701123, 5713215, 5763130, 5783316, 5787131, 5828296, 6124867, 6136125, 6150803, 6167974, 6167975, 6218358, 6222322, 6258395, 6365430, 6365431, 6390225, 6425013, 6456634, 6487035, 6556704, 6624880, 6650136, 6668870, 6743859, 6844968, 6926708, 6983799, 7008819, 7038706, 7221395, 7138319. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The CMF Battery Powered Driver graphic case contains outlines of the Battery Powered Driver, taps, screwdriver blades, and a screw holding sleeve. However, the device is not intended to be used in conjunction with a tap for tapping holes in a patient's skull, upper and/or lower jaw. Therefore some outlines in the graphic case suggest a potential use of this device that is considered off-label.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes, Inc.
- Distributor
- Nationwide Distribution
- Lot
- Part no. 305.840, with lot nos.: 5680577, 5701123, 5713215, 5763130, 5783316, 5787131, 5828296, 6124867, 6136125, 6150803, 6167974, 6167975, 6218358, 6222322, 6258395, 6365430, 6365431, 6390225, 6425013, 6456634, 6487035, 6556704, 6624880, 6650136, 6668870, 6743859, 6844968, 6926708, 6983799, 7008819, 7038706, 7221395, 7138319.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile TubeFebruary 5, 2020
- FDA2.4MM TI VA LOCKING SCREW STARDRIVE 18MM STERILE, Part Number 04.210.118TS - Product Usage:Sterile TubeFebruary 5, 2020
- FDA3.7mm Cannulated Locking Screws and 3.7mm Cannulated Conical Screws; . the Cannulated Screws are indicated forJune 24, 2015
- FDAPull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling isJune 24, 2015
- FDAGambro Cartridge Blood Set, blood transport system for hemodialysis, Model Number(s): 101025 (003410510) -April 23, 2014
- FDAThe Synthes Hohmann Retractor, 43 mm Long Shank, Angled 235 mm Product Usage: Used to expose bone forApril 23, 2014
Questions about this recall
Was Synthes CMF Battery Powered Driver, Part no. 305.840. Used to place screws in the cranialfacial bones, and to drill and/or tap holes to prepare cranialfacial bones for screw insertion. recalled?
Yes. FDA published recall Z-1456-2014 on April 23, 2014. Reported hazard: The CMF Battery Powered Driver graphic case contains outlines of the Battery Powered Driver, taps, screwdriver blades, and a screw holding sleeve. However, the device is not intended to be used in conjunction with a tap for tapping holes in a patient's skull, upper and/or lower jaw. Therefore some outlines in the graphic case suggest a potential use of this device that is considered off-label.
What should I do if I have this product?
Voluntary: Firm initiated · Letter