FDAFDA · Food and drugs·Recall number Z-1456-2014
Synthes CMF Battery Powered Driver, Part no. 305.840. Used to place screws in the cranialfacial bones, and to drill and/or tap holes to prepare cranialfacial bones for screw insertion.
Published April 23, 2014 · Recall initiated: January 30, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The CMF Battery Powered Driver graphic case contains outlines of the Battery Powered Driver, taps, screwdriver blades, and a screw holding sleeve. However, the device is not intended to be used in conjunction with a tap for tapping holes in a patient's skull, upper and/or lower jaw. Therefore some outlines in the graphic case suggest a potential use of this device that is considered off-label.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Synthes, Inc.
- Distributor
- Nationwide Distribution
- Lot
- Part no. 305.840, with lot nos.: 5680577, 5701123, 5713215, 5763130, 5783316, 5787131, 5828296, 6124867, 6136125, 6150803, 6167974, 6167975, 6218358, 6222322, 6258395, 6365430, 6365431, 6390225, 6425013, 6456634, 6487035, 6556704, 6624880, 6650136, 6668870, 6743859, 6844968, 6926708, 6983799, 7008819, 7038706, 7221395, 7138319.
- Risk classification
- Class II