FDAFDA · Food and drugs·Recall number Z-1456-2014

Synthes CMF Battery Powered Driver, Part no. 305.840. Used to place screws in the cranialfacial bones, and to drill and/or tap holes to prepare cranialfacial bones for screw insertion.

Published April 23, 2014 · Recall initiated: January 30, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The CMF Battery Powered Driver graphic case contains outlines of the Battery Powered Driver, taps, screwdriver blades, and a screw holding sleeve. However, the device is not intended to be used in conjunction with a tap for tapping holes in a patient's skull, upper and/or lower jaw. Therefore some outlines in the graphic case suggest a potential use of this device that is considered off-label.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Synthes, Inc.
Distributor
Nationwide Distribution
Lot
Part no. 305.840, with lot nos.: 5680577, 5701123, 5713215, 5763130, 5783316, 5787131, 5828296, 6124867, 6136125, 6150803, 6167974, 6167975, 6218358, 6222322, 6258395, 6365430, 6365431, 6390225, 6425013, 6456634, 6487035, 6556704, 6624880, 6650136, 6668870, 6743859, 6844968, 6926708, 6983799, 7008819, 7038706, 7221395, 7138319.
Risk classification
Class II
Area (FDA)
device
Country
United States