FDAFDA · Food and drugs·Recall number Z-1456-2014

Synthes CMF Battery Powered Driver, Part no. 305.840. Used to place screws in the cranialfacial bones, and to drill and/or tap holes to prepare cranialfacial bones for screw insertion.

Published April 23, 2014 · Recall initiated: January 30, 2014

Summary

FDA records medical device Class II recall Z-1456-2014, published April 23, 2014 and initiated January 30, 2014. Recalling firm: Synthes, Inc.; Status: Terminated; Reported quantity: 20081; Lot: Part no. 305.840, with lot nos.: 5680577, 5701123, 5713215, 5763130, 5783316, 5787131, 5828296, 6124867, 6136125, 6150803, 6167974, 6167975, 6218358, 6222322, 6258395, 6365430, 6365431, 6390225, 6425013, 6456634, 6487035, 6556704, 6624880, 6650136, 6668870, 6743859, 6844968, 6926708, 6983799, 7008819, 7038706, 7221395, 7138319. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The CMF Battery Powered Driver graphic case contains outlines of the Battery Powered Driver, taps, screwdriver blades, and a screw holding sleeve. However, the device is not intended to be used in conjunction with a tap for tapping holes in a patient's skull, upper and/or lower jaw. Therefore some outlines in the graphic case suggest a potential use of this device that is considered off-label.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Synthes, Inc.
Distributor
Nationwide Distribution
Lot
Part no. 305.840, with lot nos.: 5680577, 5701123, 5713215, 5763130, 5783316, 5787131, 5828296, 6124867, 6136125, 6150803, 6167974, 6167975, 6218358, 6222322, 6258395, 6365430, 6365431, 6390225, 6425013, 6456634, 6487035, 6556704, 6624880, 6650136, 6668870, 6743859, 6844968, 6926708, 6983799, 7008819, 7038706, 7221395, 7138319.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Synthes CMF Battery Powered Driver, Part no. 305.840. Used to place screws in the cranialfacial bones, and to drill and/or tap holes to prepare cranialfacial bones for screw insertion. recalled?

Yes. FDA published recall Z-1456-2014 on April 23, 2014. Reported hazard: The CMF Battery Powered Driver graphic case contains outlines of the Battery Powered Driver, taps, screwdriver blades, and a screw holding sleeve. However, the device is not intended to be used in conjunction with a tap for tapping holes in a patient's skull, upper and/or lower jaw. Therefore some outlines in the graphic case suggest a potential use of this device that is considered off-label.

What should I do if I have this product?

Voluntary: Firm initiated · Letter