Brilliance CT 16, Brilliance CT 64, Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology, Brilliance iCT, Ingenuity CT, Ingenuity Core, Ingenuity Core 128 and Ingenuity Flex scanners are Computed Tomography X-ray systems.
Published April 23, 2014 · Recall initiated: March 27, 2014
Summary
FDA records medical device Class II recall Z-1461-2014, published April 23, 2014 and initiated March 27, 2014. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 135; Lot: Brilliance CT 16: Model Number: 728246; Serial Number: 5314, 6719, 6726, 6727, 6728, 6729, 6730, 6731, 6732, 6734, 6735, 6736, 6737, 6738, 6744, 6746, 6747, 6748, 6749, 6750, 6751, 6752, 6753, 6754, 6756, 6757, 6758, 6759, 6760, 6761, 6762, 6764, 6767, 6768, 6772, and 50213. Brilliance CT 64: Model Number: 728231; Serial Number: 10805, 10806, 90047, 95880, 95881, 95882, 95885, 95892, 95903, 95907, 95910, 95912, 95913, 95914, 95915, 95918, 95920, 95921, 95922, and 98895. Brilliance CT Big Bore Oncology: Model Number: 728243; Serial Number: 7971, 7977, 7980, 7989, and 7994. Brilliance CT Big Bore Radiology: Model Number: 728244; Serial Number: 7957, 7965, 7979, and 7985. Brilliance iCT: Model Number: 728306; Serial Number: 100477, 100483, 100487, 100489, 100491, 100492, 100494, 100495, 100496, 100500, 100502, 100504, and 100507. Ingenuity CT: Model Number: 728326; Serial Number: 300165 and 333013. Ingenuity Core: Model Number: 728321; Serial Number: 310138, 310142, 310144, 310145, 310149, 310151, 310152, 310153, 310154, 310155, 310158, 310160, 310161, 310162, 310163, 310164, 333013, 333015, and 333017. Ingenuity Core128: Model Number: 728323; Serial Number: 320113, 320114, 320116, 320117, 320118, 320120, 320121, 320122, 320124, 320125, 320128,320129, 320130, 320134, 320137, 320138, 320139, 320140, 320141, 320142, 320143, 320144,320145, 320146, 320147, 320148, 336009, 336010, 336011, 336012, 336013, 336015, 336016, 336018, and 336023. Ingenuity Flex: Model Number: 728317; Serial Number: 345005. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Nationwide Distribution - USA including the states of CA, FL, IL, IN, KY, MD,SC, and TX.
- Lot
- Brilliance CT 16: Model Number: 728246; Serial Number: 5314, 6719, 6726, 6727, 6728, 6729, 6730, 6731, 6732, 6734, 6735, 6736, 6737, 6738, 6744, 6746, 6747, 6748, 6749, 6750, 6751, 6752, 6753, 6754, 6756, 6757, 6758, 6759, 6760, 6761, 6762, 6764, 6767, 6768, 6772, and 50213. Brilliance CT 64: Model Number: 728231; Serial Number: 10805, 10806, 90047, 95880, 95881, 95882, 95885, 95892, 95903, 95907, 95910, 95912, 95913, 95914, 95915, 95918, 95920, 95921, 95922, and 98895. Brilliance CT Big Bore Oncology: Model Number: 728243; Serial Number: 7971, 7977, 7980, 7989, and 7994. Brilliance CT Big Bore Radiology: Model Number: 728244; Serial Number: 7957, 7965, 7979, and 7985. Brilliance iCT: Model Number: 728306; Serial Number: 100477, 100483, 100487, 100489, 100491, 100492, 100494, 100495, 100496, 100500, 100502, 100504, and 100507. Ingenuity CT: Model Number: 728326; Serial Number: 300165 and 333013. Ingenuity Core: Model Number: 728321; Serial Number: 310138, 310142, 310144, 310145, 310149, 310151, 310152, 310153, 310154, 310155, 310158, 310160, 310161, 310162, 310163, 310164, 333013, 333015, and 333017. Ingenuity Core128: Model Number: 728323; Serial Number: 320113, 320114, 320116, 320117, 320118, 320120, 320121, 320122, 320124, 320125, 320128,320129, 320130, 320134, 320137, 320138, 320139, 320140, 320141, 320142, 320143, 320144,320145, 320146, 320147, 320148, 336009, 336010, 336011, 336012, 336013, 336015, 336016, 336018, and 336023. Ingenuity Flex: Model Number: 728317; Serial Number: 345005
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDAAXIOM Vertix MD Trauma systems radiographic X-rayApril 23, 2014
- FDAThe OEC 9900 Elite mobile fluoroscopy system (image intensified fluoroscopic x-ray) is designed to provideApril 23, 2014
Questions about this recall
Was Brilliance CT 16, Brilliance CT 64, Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology, Brilliance iCT, Ingenuity CT, Ingenuity Core, Ingenuity Core 128 and Ingenuity Flex scanners are Computed Tomography X-ray systems. recalled?
Yes. FDA published recall Z-1461-2014 on April 23, 2014. Reported hazard: Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.
What should I do if I have this product?
Voluntary: Firm initiated · Letter