RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system
Published March 18, 2020 · Recall initiated: February 20, 2020
Summary
FDA records medical device Class II recall Z-1463-2020, published March 18, 2020 and initiated February 20, 2020. Recalling firm: RAYSEARCH LABORATORIES AB; Status: Terminated; Reported quantity: 1 system in the U.S.; Lot: UDI 07350002010136. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Distributor
- The products were distributed to the following US states: TN
- Lot
- UDI 07350002010136
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Sweden
Related recalls
- FDARayCare 2024A SP4. Oncology Information Systems.July 8, 2026
- FDARayCare 2024A SP1. Oncology Information Systems.July 8, 2026
- FDABrand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 SoftwareJanuary 28, 2026
- FDABrand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version:January 28, 2026
- FDANextra Hammertoe Correction System-4.5MM Middle, 3.2MM Proximal, SS, Sterile. REF: NX-4532K - Product Usage:March 18, 2020
- FDARayStation/RayPlan 8B Service Pack 1, Build Number 8.1.1.8, stand-alone software treatment planning systemMarch 18, 2020
Questions about this recall
Was RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system recalled?
Yes. FDA published recall Z-1463-2020 on March 18, 2020. Reported hazard: There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail