FDAFDA · Food and drugs·Recall number Z-1463-2020
RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system
Published March 18, 2020 · Recall initiated: February 20, 2020
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Distributor
- The products were distributed to the following US states: TN
- Lot
- UDI 07350002010136
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Sweden