FDAFDA · Food and drugs·Recall number Z-1473-2013

SlingBar 350. Designed to meet the needs for lifting humans.

Published June 12, 2013 · Recall initiated: April 30, 2013

Summary

FDA records medical device Class II recall Z-1473-2013, published June 12, 2013 and initiated April 30, 2013. Recalling firm: Hill-Rom, Inc.; Status: Terminated; Reported quantity: 3556; Lot: 3156015, 3156002, 3156008' 3156004. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Hill-Rom, Inc.
Distributor
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, IL, IN, KS, LA, MA, MI, MN, MO, NC, NH, NJ, NV, OH, OK, PA, SC, TN, UT, WA, and WI, and the countries of Canada, Australia, Austria, Belgium, Switzerland, Germany Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Norway, New Zealand, Portugal, Sweden, and Turkey.
Lot
3156015, 3156002, 3156008' 3156004
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SlingBar 350. Designed to meet the needs for lifting humans. recalled?

Yes. FDA published recall Z-1473-2013 on June 12, 2013. Reported hazard: Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design.

What should I do if I have this product?

Voluntary: Firm initiated · Letter