ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.
Published April 10, 2024 · Recall initiated: February 21, 2024
Summary
FDA records medical device Class II recall Z-1476-2024, published April 10, 2024 and initiated February 21, 2024. Recalling firm: Abbott Laboratories; Status: Ongoing; Reported quantity: 7482 kits; Lot: LN 2K43-25 - Lot number 50808UN23, exp. 11/30/2024, UDI (01)00380740003302 (17)241130(10)50808UN23; LN 2K43-20 - Lot number 60104UN23, exp. 11/30/2024, UDI (01)00380740003296 (17)241130(10)60104UN23. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis of myocardial infarction.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Laboratories
- Distributor
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IL, KS, LA, MI, MO, MS, NE, NY, OH, OR, PA, TX, VA, and WI. The countries of Angola, Austria, Belarus, Belgium, Brazil, Czech Republic, France, Germany, Hungary, India, Ireland, Israel, Italy, Japan, Latvia, Lithuania, Luxembourg, Mali, Mexico, Morocco, Poland, Portugal, Peoples Republic of China, Romania, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Switzerland, Taiwan, Tajikistan, Tanzania, Trinidad & Tobago, Turkey, Uganda, Uruguay, Uzbekistan, and Vietnam.
- Lot
- LN 2K43-25 - Lot number 50808UN23, exp. 11/30/2024, UDI (01)00380740003302 (17)241130(10)50808UN23; LN 2K43-20 - Lot number 60104UN23, exp. 11/30/2024, UDI (01)00380740003296 (17)241130(10)60104UN23.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAlinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20December 24, 2025
- FDABrand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: AllApril 30, 2025
- FDAAlinity hq Analyzer, REF: 09P68-01August 28, 2024
- FDA3M Tube Securement Device, Small, 1.7 IN/PO x 2.4 IN/PO, REF 1500U, Sterile; and 3M Tube Securement Device,April 10, 2024
- FDAstryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplastyApril 10, 2024
- FDAARCHITECT c4000 Processing Module REF: 02P24-01 02P24-02 02P24-40 01R24-56 01R25-56 The Abbott ARCHITECT c4000January 17, 2024
Questions about this recall
Was ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20. recalled?
Yes. FDA published recall Z-1476-2024 on April 10, 2024. Reported hazard: The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis of myocardial infarction.
What should I do if I have this product?
Voluntary: Firm initiated · Letter