Alinity hq Analyzer, REF: 09P68-01

Published August 28, 2024 · Recall initiated: June 14, 2024

Summary

FDA records medical device Class II recall Z-2681-2024, published August 28, 2024 and initiated June 14, 2024. Recalling firm: Abbott Laboratories; Status: Ongoing; Reported quantity: 816; Lot: UDI-DI: 380740138851. Software Versions: 5.7 (US) and 5.6 and below (OUS), Serial Numbers: HQ00105, HQ00108, HQ00117, HQ00118, HQ00119, HQ00120, HQ00126, HQ00127, HQ00128, HQ00132, HQ00136, HQ00139, HQ00140, HQ00141, HQ00142, HQ00144, HQ00145, HQ00147, HQ00148, HQ00149, HQ00152, HQ00153, HQ00154, HQ00156, HQ00157, HQ00161, HQ00162, HQ00166, HQ00169, HQ00171, HQ00172, HQ00173, HQ00174, HQ00175, HQ00183, HQ00184, HQ00185, HQ00186, HQ00187, HQ00188, HQ00266, HQ00268, HQ00269, HQ00270, HQ00271, HQ00273, HQ00274, HQ00276, HQ00277, HQ00278, HQ00279, HQ00281, HQ00282, HQ00283, HQ00286, HQ00287, HQ00289, HQ00291, HQ00295, HQ00296, HQ00299, HQ00300, HQ00302, HQ00303, HQ00312, HQ00313, HQ00314, HQ00316, HQ00318, HQ00319, HQ00321, HQ00324, HQ00327, HQ00328, HQ00329, HQ00331, HQ00332, HQ00334, HQ00335, HQ00336, HQ00337, HQ00338, HQ00417, HQ00418, HQ00419, HQ00420, HQ00422, HQ00424, HQ00425, HQ00430, HQ00432, HQ00433, HQ00434, HQ00435, HQ00436, HQ00438, HQ00439, HQ00440, HQ00441, HQ00442, HQ00443, HQ00444, HQ00445, HQ00446, HQ00447, HQ00448, HQ00451, HQ00453, HQ00455, HQ00456, HQ00457, HQ00458, HQ00460, HQ00461, HQ00463, HQ00464, HQ00466, HQ00468, HQ00469, HQ00471, HQ00472, HQ00474, HQ00475, HQ00476, HQ00536, HQ00539, HQ00540, HQ00541, HQ00542, HQ00544, HQ00545, HQ00546, HQ00547, HQ00548, HQ00550, HQ00551, HQ00552, HQ00553, HQ00554, HQ00555, HQ00558, HQ00561, HQ00562, HQ00563, HQ00564, HQ00565, HQ00567, HQ00568, HQ00569, HQ00570, HQ00571, HQ00573, HQ00574, HQ00575, HQ00576, HQ00577, HQ00578, HQ00579, HQ00580, HQ00581, HQ00582, HQ00583, HQ00584, HQ00585, HQ00586, HQ00587, HQ00638, HQ00639, HQ00640, HQ00641, HQ00642, HQ00643, HQ00644, HQ00645, HQ00646, HQ00647, HQ00648, HQ00649, HQ00650, HQ00651, HQ00654, HQ00655, HQ00657, HQ00660, HQ00662, HQ00663, HQ00665, HQ00666, HQ00667, HQ00668, HQ00669, HQ00670, HQ00671, HQ00672, HQ00673, HQ00674, HQ00675, HQ00676, HQ00677, HQ00678, HQ00679, HQ00680, HQ00681, HQ00682, HQ00683, HQ00684, HQ00685, HQ00686, HQ00731, HQ00732, HQ00733, HQ00734, HQ00735, HQ00736, HQ00737, HQ00738, HQ00740, HQ00741, HQ00742, HQ00743, HQ00744, HQ00745, HQ00746, HQ00747, HQ00748, HQ00749, HQ00750, HQ00751, HQ00753, HQ00754, HQ00756, HQ00757, HQ00758, HQ00759, HQ00760, HQ00761, HQ00762, HQ00763, HQ00764, HQ00765, HQ00766, HQ00767, HQ00768, HQ00769, HQ00770, HQ00771, HQ00772, HQ00773, HQ00774, HQ00775, HQ00819, HQ00820, HQ00821, HQ00822, HQ00823, HQ00824, HQ00825, HQ00826, HQ00827, HQ00828, HQ00829, HQ00830, HQ00831, HQ00832, HQ00833, HQ00834, HQ00835, HQ00836, HQ00837, HQ00838, HQ00839, HQ00840, HQ00841, HQ00842, HQ00844, HQ00845, HQ00846, HQ00847, HQ00848, HQ00849, HQ00852, HQ00853, HQ00856, HQ00857, HQ00858, HQ00859, HQ00860, HQ00861, HQ00864, HQ00865, HQ00866, HQ00867, HQ00923, HQ00924, HQ00925, HQ00926, HQ00927, HQ00929, HQ00932, HQ00933, HQ00934, HQ00935, HQ00936, HQ00937, HQ00938, HQ00939, HQ00940, HQ00941, HQ00942, HQ00943, HQ00944, HQ00945, HQ00946, HQ00947, HQ00949, HQ00950, HQ00951, HQ00952, HQ00953, HQ00954, HQ00955, HQ00956, HQ00957, HQ00958, HQ00959, HQ00960, HQ00961, HQ00962, HQ00963, HQ00964, HQ00965, HQ00966, HQ00967, HQ00968, HQ01018, HQ01019, HQ01020, HQ01021, HQ01022, HQ01023, HQ01024, HQ01025, HQ01026, HQ01027, HQ01028, HQ01029, HQ01030, HQ01031, HQ01032, HQ01033, HQ01034, HQ01035, HQ01037, HQ01038, HQ01039, HQ01040, HQ01043, HQ01044, HQ01046, HQ01047, HQ01049, HQ01050, HQ01051, HQ01052, HQ01054, HQ01055, HQ01056, HQ01057, HQ01058, HQ01060, HQ01061, HQ01062, HQ01063, HQ01064, HQ01065, HQ01067, HQ01122, HQ01123, HQ01124, HQ01125, HQ01126, HQ01127, HQ01128, HQ01129, HQ01130, HQ01131, HQ01132, HQ01133, HQ01134, HQ01135, HQ01136, HQ01137, HQ01138, HQ01139, HQ01140, HQ01141, HQ01142, HQ01143, HQ01144, HQ01145, HQ01146, HQ01147, HQ01148, HQ01149, HQ01150, HQ01151, HQ01152, HQ01153, HQ00191, HQ00192, HQ00193, HQ00194, HQ00196, HQ00197, HQ00198, HQ00201, HQ00202, HQ00203, HQ00204, HQ00208, HQ00209, HQ00210, HQ00214, HQ00215, HQ00217, HQ00218, HQ00222, HQ00223, HQ00230, HQ00231, HQ00233, HQ00242, HQ00243, HQ00244, HQ. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell Hemoglobin(MCH)/MCHC and falsely low hematocrit results. 2)Cell events may be incorrectly counted as basophil(BASO), resulting in increased BASO and %BASO counts

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Abbott Laboratories
Distributor
Worldwide - US Nationwide distribution in the states of NY, OK, NC and the countries of ALBANIA, ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, Botswana, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAMAICA, JAPAN, JORDAN, KUWAIT, LEBANON, LITHUANIA, MALAYSIA, MALDIVES, MEXICO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SERBIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, URUGUAY, VIETNAM.
Lot
UDI-DI: 380740138851. Software Versions: 5.7 (US) and 5.6 and below (OUS), Serial Numbers: HQ00105, HQ00108, HQ00117, HQ00118, HQ00119, HQ00120, HQ00126, HQ00127, HQ00128, HQ00132, HQ00136, HQ00139, HQ00140, HQ00141, HQ00142, HQ00144, HQ00145, HQ00147, HQ00148, HQ00149, HQ00152, HQ00153, HQ00154, HQ00156, HQ00157, HQ00161, HQ00162, HQ00166, HQ00169, HQ00171, HQ00172, HQ00173, HQ00174, HQ00175, HQ00183, HQ00184, HQ00185, HQ00186, HQ00187, HQ00188, HQ00266, HQ00268, HQ00269, HQ00270, HQ00271, HQ00273, HQ00274, HQ00276, HQ00277, HQ00278, HQ00279, HQ00281, HQ00282, HQ00283, HQ00286, HQ00287, HQ00289, HQ00291, HQ00295, HQ00296, HQ00299, HQ00300, HQ00302, HQ00303, HQ00312, HQ00313, HQ00314, HQ00316, HQ00318, HQ00319, HQ00321, HQ00324, HQ00327, HQ00328, HQ00329, HQ00331, HQ00332, HQ00334, HQ00335, HQ00336, HQ00337, HQ00338, HQ00417, HQ00418, HQ00419, HQ00420, HQ00422, HQ00424, HQ00425, HQ00430, HQ00432, HQ00433, HQ00434, HQ00435, HQ00436, HQ00438, HQ00439, HQ00440, HQ00441, HQ00442, HQ00443, HQ00444, HQ00445, HQ00446, HQ00447, HQ00448, HQ00451, HQ00453, HQ00455, HQ00456, HQ00457, HQ00458, HQ00460, HQ00461, HQ00463, HQ00464, HQ00466, HQ00468, HQ00469, HQ00471, HQ00472, HQ00474, HQ00475, HQ00476, HQ00536, HQ00539, HQ00540, HQ00541, HQ00542, HQ00544, HQ00545, HQ00546, HQ00547, HQ00548, HQ00550, HQ00551, HQ00552, HQ00553, HQ00554, HQ00555, HQ00558, HQ00561, HQ00562, HQ00563, HQ00564, HQ00565, HQ00567, HQ00568, HQ00569, HQ00570, HQ00571, HQ00573, HQ00574, HQ00575, HQ00576, HQ00577, HQ00578, HQ00579, HQ00580, HQ00581, HQ00582, HQ00583, HQ00584, HQ00585, HQ00586, HQ00587, HQ00638, HQ00639, HQ00640, HQ00641, HQ00642, HQ00643, HQ00644, HQ00645, HQ00646, HQ00647, HQ00648, HQ00649, HQ00650, HQ00651, HQ00654, HQ00655, HQ00657, HQ00660, HQ00662, HQ00663, HQ00665, HQ00666, HQ00667, HQ00668, HQ00669, HQ00670, HQ00671, HQ00672, HQ00673, HQ00674, HQ00675, HQ00676, HQ00677, HQ00678, HQ00679, HQ00680, HQ00681, HQ00682, HQ00683, HQ00684, HQ00685, HQ00686, HQ00731, HQ00732, HQ00733, HQ00734, HQ00735, HQ00736, HQ00737, HQ00738, HQ00740, HQ00741, HQ00742, HQ00743, HQ00744, HQ00745, HQ00746, HQ00747, HQ00748, HQ00749, HQ00750, HQ00751, HQ00753, HQ00754, HQ00756, HQ00757, HQ00758, HQ00759, HQ00760, HQ00761, HQ00762, HQ00763, HQ00764, HQ00765, HQ00766, HQ00767, HQ00768, HQ00769, HQ00770, HQ00771, HQ00772, HQ00773, HQ00774, HQ00775, HQ00819, HQ00820, HQ00821, HQ00822, HQ00823, HQ00824, HQ00825, HQ00826, HQ00827, HQ00828, HQ00829, HQ00830, HQ00831, HQ00832, HQ00833, HQ00834, HQ00835, HQ00836, HQ00837, HQ00838, HQ00839, HQ00840, HQ00841, HQ00842, HQ00844, HQ00845, HQ00846, HQ00847, HQ00848, HQ00849, HQ00852, HQ00853, HQ00856, HQ00857, HQ00858, HQ00859, HQ00860, HQ00861, HQ00864, HQ00865, HQ00866, HQ00867, HQ00923, HQ00924, HQ00925, HQ00926, HQ00927, HQ00929, HQ00932, HQ00933, HQ00934, HQ00935, HQ00936, HQ00937, HQ00938, HQ00939, HQ00940, HQ00941, HQ00942, HQ00943, HQ00944, HQ00945, HQ00946, HQ00947, HQ00949, HQ00950, HQ00951, HQ00952, HQ00953, HQ00954, HQ00955, HQ00956, HQ00957, HQ00958, HQ00959, HQ00960, HQ00961, HQ00962, HQ00963, HQ00964, HQ00965, HQ00966, HQ00967, HQ00968, HQ01018, HQ01019, HQ01020, HQ01021, HQ01022, HQ01023, HQ01024, HQ01025, HQ01026, HQ01027, HQ01028, HQ01029, HQ01030, HQ01031, HQ01032, HQ01033, HQ01034, HQ01035, HQ01037, HQ01038, HQ01039, HQ01040, HQ01043, HQ01044, HQ01046, HQ01047, HQ01049, HQ01050, HQ01051, HQ01052, HQ01054, HQ01055, HQ01056, HQ01057, HQ01058, HQ01060, HQ01061, HQ01062, HQ01063, HQ01064, HQ01065, HQ01067, HQ01122, HQ01123, HQ01124, HQ01125, HQ01126, HQ01127, HQ01128, HQ01129, HQ01130, HQ01131, HQ01132, HQ01133, HQ01134, HQ01135, HQ01136, HQ01137, HQ01138, HQ01139, HQ01140, HQ01141, HQ01142, HQ01143, HQ01144, HQ01145, HQ01146, HQ01147, HQ01148, HQ01149, HQ01150, HQ01151, HQ01152, HQ01153, HQ00191, HQ00192, HQ00193, HQ00194, HQ00196, HQ00197, HQ00198, HQ00201, HQ00202, HQ00203, HQ00204, HQ00208, HQ00209, HQ00210, HQ00214, HQ00215, HQ00217, HQ00218, HQ00222, HQ00223, HQ00230, HQ00231, HQ00233, HQ00242, HQ00243, HQ00244, HQ
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Alinity hq Analyzer, REF: 09P68-01 recalled?

Yes. FDA published recall Z-2681-2024 on August 28, 2024. Reported hazard: 1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell Hemoglobin(MCH)/MCHC and falsely low hematocrit results. 2)Cell events may be incorrectly counted as basophil(BASO), resulting in increased BASO and %BASO counts

What should I do if I have this product?

Voluntary: Firm initiated · Letter