C2 CryoBalloon Ablation System - Controller Cap. C2 CryoBalloon Controller; Cryosurgical Unit, Cryogenic Surgical Device The affected product is the Controller Cap component only, which is provided to the customer already assembled onto the Controller. Model number: FG-1012 Product Usage: The C2 CryoBalloon Ablation System is intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett s Esophagus with dysplasia. It is a cryosurgical system with a nitrous oxide cooled balloon that is compatible with a commercially available endoscope with a minimum working channel inner diameter (ID) of 3.7 mm and length of 100 cm. The C2 CryoBalloon System is a system comprised of a Catheter (sterile), Controller (non-sterile), and Cartridge (non-sterile). The Device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the inflated balloon, and the treatment site is selected by adjusting the endoscope and Controller position. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Controller.
Published March 22, 2017 · Recall initiated: January 6, 2017
Summary
FDA records medical device Class II recall Z-1477-2017, published March 22, 2017 and initiated January 6, 2017. Recalling firm: C2 Therapeutics, Inc.; Status: Terminated; Reported quantity: 755 controller caps; Lot: All lots distributed between June 23, 2016 and December 20, 2016: UPDATE - expanded to include product distributed from February 1 2016 through February 8, 2017. All numbers below have been updated. Lot numbers: Expiry dates 01122016-02 12-Jul-16; 06242016-01 1-Jan-2017; 11282016-03 2-Jun-2017; 01042016-03 14-Jul-16; 06292016-02 6-Jan-2017; 09202016-01 21-Mar-2017; 07062016-01 8-Jan-2017; 09222016-01 23-Mar-2017; 01252016-02 02-Aug-16; 07082016-02 13-Jan 2017; 09232016-01 28-Mar-2017; 01282016-01 01-Aug-16; 07082016-03 14-Jan-2017; 09232016-02 27-Mar-2017; 01292016-01 02-Aug-16; 07152016-01 21-Jan-2017; 09292016-01 4-Apr-2017; 02022016-03 09-Aug-16; 07152016-02 20-Jan-2017; 09302016-03 4-Apr-2017; 02082016-01 09-Aug-16; 07212016-01 25-Jan-2017; 10052016-02 19-Apr-2017; 02092016-02 11-Aug-16; 07252016-01 28-Jan-2017; 10142016-01 18-Apr-2017; 02112016-02 17-Aug-16; 07262016-01 9-Feb-2017; 10192016-01 24-Apr-2017; 02172016-02 25-Aug-16; 07282016-02 2-Feb-2017; 10212016-01 25-Apr-2017; 02222016-01 26-Aug-16; 08022016-01 17-Feb-2017; 10242016-01 26-Apr-2017; 03012016-01 04-Sep-16; 08032016-02 5-Feb-2017; 10262016-02 1-May-2017; 03072016-01 10-Sep-16; 08112016-01 16-Feb-2017; 10282016-02 8-May-2017; 03102016-02 15-Sep-16; 08232016-01 26-Feb-2017; 11022016-01 9-May-2017; 03152016-03 21-Sep-16; 08262016-01 2-Mar-2017; 11042016-01 10-May-2017; 03282016-02 29-Sep-16; 09022016-01 8-Mar-2017; 11092016-01 15-May-2017; 03292016-01 31-Sep-16; 09092016-01 14-Mar-2017; 11142016-02 16-May-2017; 09122016-02; 15-Mar-2017; 11152016-02 18-May-2017; 04012016-01 06-Oct-16; 09192016-01 19-Mar-2017; 11162016-01 21-May-2017; 09192016-02 21-Mar-2017; 11282016-02 30-May-2017; 04062016-01 12-Oct-16; 11302016-02 02-Jun-17; 12052016-01 08-Jun-17; 04112016-04 14-Oct-16; 12052016-03 06-Jun-17; 04142016-01 20-Oct-16; 12072016-01 09-Jun-17; 04202016-02 21-Oct-16; 12092016-01 13-Jun-17; 05022016-02 05-Nov-16; 12132016-01 19-Jun-17; 05092016-01 11-Nov-16; 12152016-02 20-Jun-17; 05162016-02 18-Nov-16; 01042017-01 04-Jul-17; 05232016-01 25-Nov-16; 01052017-03 19-Jul-17; 05272016-01 02-Dec-16; 01052017-04 19-Jul-17; 06022016-02 09-Dec-16; 01092017-01 12-Jul-17; 06102016-01 14-Dec-16; 01102017-01 18-Jul-17; 06152016-01 21-Dec-16; 01132017-01 18-Jul-17; 06172016-01 23-Dec-16; 01132017-03 17-Jul-17; 07282016-02 02-Feb-17. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Controller Cap of the C2 CyroBalloon Ablation System may crack as it is tightened onto the controller. That could compromise the integrity and result in damage to the Controller Cap.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- C2 Therapeutics, Inc.
- Distributor
- Worldwide Distribution - US Nationwide and the countries of Germany, Belgium, The Netherlands United Kingdom and China.
- Lot
- All lots distributed between June 23, 2016 and December 20, 2016: UPDATE - expanded to include product distributed from February 1 2016 through February 8, 2017. All numbers below have been updated. Lot numbers: Expiry dates 01122016-02 12-Jul-16; 06242016-01 1-Jan-2017; 11282016-03 2-Jun-2017; 01042016-03 14-Jul-16; 06292016-02 6-Jan-2017; 09202016-01 21-Mar-2017; 07062016-01 8-Jan-2017; 09222016-01 23-Mar-2017; 01252016-02 02-Aug-16; 07082016-02 13-Jan 2017; 09232016-01 28-Mar-2017; 01282016-01 01-Aug-16; 07082016-03 14-Jan-2017; 09232016-02 27-Mar-2017; 01292016-01 02-Aug-16; 07152016-01 21-Jan-2017; 09292016-01 4-Apr-2017; 02022016-03 09-Aug-16; 07152016-02 20-Jan-2017; 09302016-03 4-Apr-2017; 02082016-01 09-Aug-16; 07212016-01 25-Jan-2017; 10052016-02 19-Apr-2017; 02092016-02 11-Aug-16; 07252016-01 28-Jan-2017; 10142016-01 18-Apr-2017; 02112016-02 17-Aug-16; 07262016-01 9-Feb-2017; 10192016-01 24-Apr-2017; 02172016-02 25-Aug-16; 07282016-02 2-Feb-2017; 10212016-01 25-Apr-2017; 02222016-01 26-Aug-16; 08022016-01 17-Feb-2017; 10242016-01 26-Apr-2017; 03012016-01 04-Sep-16; 08032016-02 5-Feb-2017; 10262016-02 1-May-2017; 03072016-01 10-Sep-16; 08112016-01 16-Feb-2017; 10282016-02 8-May-2017; 03102016-02 15-Sep-16; 08232016-01 26-Feb-2017; 11022016-01 9-May-2017; 03152016-03 21-Sep-16; 08262016-01 2-Mar-2017; 11042016-01 10-May-2017; 03282016-02 29-Sep-16; 09022016-01 8-Mar-2017; 11092016-01 15-May-2017; 03292016-01 31-Sep-16; 09092016-01 14-Mar-2017; 11142016-02 16-May-2017; 09122016-02; 15-Mar-2017; 11152016-02 18-May-2017; 04012016-01 06-Oct-16; 09192016-01 19-Mar-2017; 11162016-01 21-May-2017; 09192016-02 21-Mar-2017; 11282016-02 30-May-2017; 04062016-01 12-Oct-16; 11302016-02 02-Jun-17; 12052016-01 08-Jun-17; 04112016-04 14-Oct-16; 12052016-03 06-Jun-17; 04142016-01 20-Oct-16; 12072016-01 09-Jun-17; 04202016-02 21-Oct-16; 12092016-01 13-Jun-17; 05022016-02 05-Nov-16; 12132016-01 19-Jun-17; 05092016-01 11-Nov-16; 12152016-02 20-Jun-17; 05162016-02 18-Nov-16; 01042017-01 04-Jul-17; 05232016-01 25-Nov-16; 01052017-03 19-Jul-17; 05272016-01 02-Dec-16; 01052017-04 19-Jul-17; 06022016-02 09-Dec-16; 01092017-01 12-Jul-17; 06102016-01 14-Dec-16; 01102017-01 18-Jul-17; 06152016-01 21-Dec-16; 01132017-01 18-Jul-17; 06172016-01 23-Dec-16; 01132017-03 17-Jul-17; 07282016-02 02-Feb-17.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAA+D Diaper Rash Cream with Dimethicone 1% and Zinc Oxide 10%, Packaged in a) 1.5 oz (42.5g) ( UPCMarch 29, 2017
- FDAVITROS Chemistry Products Li (Lithium) Slides, REF/Catalog Number/Product Code 1632660 (18 slide format, 90March 22, 2017
- FDAInion GRT Tack, drill bit and tack appl., 4x, Sterile. Is indicated for the surgical treatment of periodontalMarch 22, 2017
- FDAArtis Q and Q.Zen fluoroscopic x-ray systemMarch 22, 2017
- FDA882480: BrightView designed for single or dual detector nuclear imaging accommodating a range of ECT studies.March 22, 2017
- FDABivona® TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.March 22, 2017
Questions about this recall
Was C2 CryoBalloon Ablation System - Controller Cap. C2 CryoBalloon Controller; Cryosurgical Unit, Cryogenic Surgical Device The affected product is the Controller Cap component only, which is provided to the customer already assembled onto the Controller. Model number: FG-1012 Product Usage: The C2 CryoBalloon Ablation System is intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett s Esophagus with dysplasia. It is a cryosurgical system with a nitrous oxide cooled balloon that is compatible with a commercially available endoscope with a minimum working channel inner diameter (ID) of 3.7 mm and length of 100 cm. The C2 CryoBalloon System is a system comprised of a Catheter (sterile), Controller (non-sterile), and Cartridge (non-sterile). The Device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the inflated balloon, and the treatment site is selected by adjusting the endoscope and Controller position. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Controller. recalled?
Yes. FDA published recall Z-1477-2017 on March 22, 2017. Reported hazard: The Controller Cap of the C2 CyroBalloon Ablation System may crack as it is tightened onto the controller. That could compromise the integrity and result in damage to the Controller Cap.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit