882480: BrightView designed for single or dual detector nuclear imaging accommodating a range of ECT studies. In addition it can be used to perform planar static, dynamic, gated, total body, circular-orbit and noncircular orbit SPECT, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors).

Published March 22, 2017 · Recall initiated: January 31, 2017

Summary

FDA records medical device Class II recall Z-1480-2017, published March 22, 2017 and initiated January 31, 2017. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 1218 total; Lot: 4000201, 40007030003, 40007060006, 4000586, 4000327, 4000307, 4000361, 4000146, 4000159, 4000077, 4000050, 4000687, 4000325, 4000165, 4000252, 4000315, 4000191, 4000598, 4000141, 4000565, 4000135, 4000537, 4000591, 4000311, 4000626, 4000636, 4000329, 4000181, 4000644, 4000171, 4000120, 4000484A, 4000441, 4000054, 4000110, 4000653, 4000590, 4000100, 4000520, 4000323, 4000403, 4000406, 4000489, 4000277, 4000306, 4000332, 4000635, 4000270, 4000302, 4000046, 4000214, 4000615, 4000187, 4000660, 4000383, 4000633, 4000322, 4000212, 4000628, 4000034, 4000232, 4000668, 4000480, 4000430, 4000553, 4000467, 4000221, 4000127, 4000259, 4000670, 4000193, 4000222, 4000289, 4000597, 4000439, 4000542, 4000531, 4000588, 4000358, 4000330, 4000508, 4000310, 4000371, 4000501, 4000609, 4000616, 4000116, 4000452, 4000058, 4000521, 4000059, 4000064, 4000482, 4000384, 4000528, 4000177, 4000506, 4000278, 4000089, 4000139, 4000130, 4000564, 4000578, 4000674, 4000083, 4000583, 4000225, 4000267, 4000563, 4000587, 4000117, 4000053, 4000662, 4000526, 4000207, 4000118, 4000103, 4000556, 4000247, 4000326, 4000174, 4000248, 4000642, 4000696, 4000182, 4000456, 4000060, 4000522, 4000448, 4000099, 4000246, 4000175, 4000524, 4000630, 4000155, 4000367, 4000360, 4000492, 4000037, 4000230, 4000308, 4000236, 4000143, 4000285, 4000505, 4000206, 4000684, 4000164, 4000004, 4000650, 4000671, 4000473, 4000405, 4000235, 4000242, 4000680, 4000318, 4000561, 4000149, 4000073, 4000240, 4000606, 4000610, 4000304, 4000584, 4000580, 4000460, 4000263, 4000347, 4000632, 4000265, 4000368, 4000055, 4000349, 4000658, 4000351, 4000388, 4000589, 4000558, 4000106, 4000144, 4000444, 4000292, 4000098A, 4000319, 4000245, 4000295, 400054, 4000470, 4000666, 4000111, 4000148, 4000076, 4000443, 4000575, 4000654, 4000251, 4000108, 4000657, 4000659, 4000562, 4000535, 4000435, 4000357,4000677, 4000453, 4000377, 4000387, 4000446, 4000608, 40007060008, 4000036, 4000433, 4000538, 4000474, 4000414, 4000569, 4000266, 4000204, 4000119, 4000079, 4000234, 4000253, 4000192, 4000249, 4000634, 4000152, 4000380, 4000513, 4000198, 4000313, 4000340, 4000525, 4000541, 4000231, 4000424, 4000161, 4000170, 4000048, 4000239, 4000124, 4000200, 40007050005, 40007060007, 4000625, 4000264, 4000534, 4000261, 4000359, 4000220, 4000136, 4000140, 4000355, 4000087, 4000104, 4000611, 4000002, 4000040, 4000093, 4000258, 4000352, 4000291, 4000132, 4000334, 4000486, 4000271, 4000599, 4000605, 4000196, 4000363, 4000656, 4000362, 4000074, 4000577, 4000151-1, 4000238, 4000007, 4000113, 4000284, 4000260, 4000244, 4000391, 4000344, 4000695, 4000112, 4000205, 4000250, 4000052, 4000432, 4000275, 4000185, 4000646, 4000094, 4000215, 4000507, 4000090, 40000101, 4000582, 4000594, 4000128, 4000661, 4000431, 4000703001, 4000523, 4000078, 4000425, 4000062, 4000145, 4000011, 4000426, 4000440, 4000529, 4000450, 4000509, 4000241, 4000320, 40000092, 4000603, 4000496, 4000532, 4000502, 4000255, 4000593, 4000490, 4000293, 4000434, 4000619, 4000438, 4000095, 4000499, 4000623, 4000517, 4000624, 4000162, 4000402, 4000574, 4000476, 4000629, 4000649, 4000685, 4000596, 4000160, 4000546, 4000622, 4000186, 4000189, 4000445, 4000429, 4000274, 4000697, 4000673, 4000038, 4000176, 4000404, 4000114, 4000309, 4000138, 4000188, 4000369, 4000178, 4000459, 4000461, 4000468, 4000063, 4000150, 4000210, 4000217, 4000547, 4000527, 4000279, 4000208, 4000216, 4000269, 4000125, 4000254, 4000345, 4000415, 4000282, 4000449, 4000679, 4000199, 4000211, 4000331, 4000082, 4000134, 4000163, 4000373, 4000409, 4000465, 4000471, 4000545, 4000618, 4000637, 4000641, 4000153, 4000539, 4000663, 4000676, 4000477, 4000299, 4000133, 4000675, 4000314, 4000686, 4000209, 4000010, 4000621, 4000041, 4000167, 4000203, 4000407, 4000543, 4000374, 4000428, 4000157, 4000013, 4000075, 4000080, 4000086, 4000392, 4000396, 4000481, 4000689, 4000420, 4000690, 4000530, 4000639, 4000479, 4000568, 4000300, 4000572, 4000573, 4000640, 4000183, 4000. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide distribution. US nationwide, ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CAYMEN ISLANDS, CHILE, CHINA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, EGYPT, FINLAND, FRANCE, GABON, GERMANY, GREAT BRITAIN, GREECE, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, LEBANON, LITHUANIA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PALESTINE, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, THAILAND, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, & VIETNAM.
Lot
4000201, 40007030003, 40007060006, 4000586, 4000327, 4000307, 4000361, 4000146, 4000159, 4000077, 4000050, 4000687, 4000325, 4000165, 4000252, 4000315, 4000191, 4000598, 4000141, 4000565, 4000135, 4000537, 4000591, 4000311, 4000626, 4000636, 4000329, 4000181, 4000644, 4000171, 4000120, 4000484A, 4000441, 4000054, 4000110, 4000653, 4000590, 4000100, 4000520, 4000323, 4000403, 4000406, 4000489, 4000277, 4000306, 4000332, 4000635, 4000270, 4000302, 4000046, 4000214, 4000615, 4000187, 4000660, 4000383, 4000633, 4000322, 4000212, 4000628, 4000034, 4000232, 4000668, 4000480, 4000430, 4000553, 4000467, 4000221, 4000127, 4000259, 4000670, 4000193, 4000222, 4000289, 4000597, 4000439, 4000542, 4000531, 4000588, 4000358, 4000330, 4000508, 4000310, 4000371, 4000501, 4000609, 4000616, 4000116, 4000452, 4000058, 4000521, 4000059, 4000064, 4000482, 4000384, 4000528, 4000177, 4000506, 4000278, 4000089, 4000139, 4000130, 4000564, 4000578, 4000674, 4000083, 4000583, 4000225, 4000267, 4000563, 4000587, 4000117, 4000053, 4000662, 4000526, 4000207, 4000118, 4000103, 4000556, 4000247, 4000326, 4000174, 4000248, 4000642, 4000696, 4000182, 4000456, 4000060, 4000522, 4000448, 4000099, 4000246, 4000175, 4000524, 4000630, 4000155, 4000367, 4000360, 4000492, 4000037, 4000230, 4000308, 4000236, 4000143, 4000285, 4000505, 4000206, 4000684, 4000164, 4000004, 4000650, 4000671, 4000473, 4000405, 4000235, 4000242, 4000680, 4000318, 4000561, 4000149, 4000073, 4000240, 4000606, 4000610, 4000304, 4000584, 4000580, 4000460, 4000263, 4000347, 4000632, 4000265, 4000368, 4000055, 4000349, 4000658, 4000351, 4000388, 4000589, 4000558, 4000106, 4000144, 4000444, 4000292, 4000098A, 4000319, 4000245, 4000295, 400054, 4000470, 4000666, 4000111, 4000148, 4000076, 4000443, 4000575, 4000654, 4000251, 4000108, 4000657, 4000659, 4000562, 4000535, 4000435, 4000357,4000677, 4000453, 4000377, 4000387, 4000446, 4000608, 40007060008, 4000036, 4000433, 4000538, 4000474, 4000414, 4000569, 4000266, 4000204, 4000119, 4000079, 4000234, 4000253, 4000192, 4000249, 4000634, 4000152, 4000380, 4000513, 4000198, 4000313, 4000340, 4000525, 4000541, 4000231, 4000424, 4000161, 4000170, 4000048, 4000239, 4000124, 4000200, 40007050005, 40007060007, 4000625, 4000264, 4000534, 4000261, 4000359, 4000220, 4000136, 4000140, 4000355, 4000087, 4000104, 4000611, 4000002, 4000040, 4000093, 4000258, 4000352, 4000291, 4000132, 4000334, 4000486, 4000271, 4000599, 4000605, 4000196, 4000363, 4000656, 4000362, 4000074, 4000577, 4000151-1, 4000238, 4000007, 4000113, 4000284, 4000260, 4000244, 4000391, 4000344, 4000695, 4000112, 4000205, 4000250, 4000052, 4000432, 4000275, 4000185, 4000646, 4000094, 4000215, 4000507, 4000090, 40000101, 4000582, 4000594, 4000128, 4000661, 4000431, 4000703001, 4000523, 4000078, 4000425, 4000062, 4000145, 4000011, 4000426, 4000440, 4000529, 4000450, 4000509, 4000241, 4000320, 40000092, 4000603, 4000496, 4000532, 4000502, 4000255, 4000593, 4000490, 4000293, 4000434, 4000619, 4000438, 4000095, 4000499, 4000623, 4000517, 4000624, 4000162, 4000402, 4000574, 4000476, 4000629, 4000649, 4000685, 4000596, 4000160, 4000546, 4000622, 4000186, 4000189, 4000445, 4000429, 4000274, 4000697, 4000673, 4000038, 4000176, 4000404, 4000114, 4000309, 4000138, 4000188, 4000369, 4000178, 4000459, 4000461, 4000468, 4000063, 4000150, 4000210, 4000217, 4000547, 4000527, 4000279, 4000208, 4000216, 4000269, 4000125, 4000254, 4000345, 4000415, 4000282, 4000449, 4000679, 4000199, 4000211, 4000331, 4000082, 4000134, 4000163, 4000373, 4000409, 4000465, 4000471, 4000545, 4000618, 4000637, 4000641, 4000153, 4000539, 4000663, 4000676, 4000477, 4000299, 4000133, 4000675, 4000314, 4000686, 4000209, 4000010, 4000621, 4000041, 4000167, 4000203, 4000407, 4000543, 4000374, 4000428, 4000157, 4000013, 4000075, 4000080, 4000086, 4000392, 4000396, 4000481, 4000689, 4000420, 4000690, 4000530, 4000639, 4000479, 4000568, 4000300, 4000572, 4000573, 4000640, 4000183, 4000
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was 882480: BrightView designed for single or dual detector nuclear imaging accommodating a range of ECT studies. In addition it can be used to perform planar static, dynamic, gated, total body, circular-orbit and noncircular orbit SPECT, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors). recalled?

Yes. FDA published recall Z-1480-2017 on March 22, 2017. Reported hazard: Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan

What should I do if I have this product?

Voluntary: Firm initiated · Letter