Merit Laureate 0.035" (0.89mm) Hydrophilic Guide Wire, Angled Tip, Standard Shaft, Catalog No. LWSTDA35260EX, 260 cm (102").
Published April 30, 2014 · Recall initiated: April 14, 2014
Summary
FDA records medical device Class II recall Z-1486-2014, published April 30, 2014 and initiated April 14, 2014. Recalling firm: Merit Medical Systems, Inc.; Status: Terminated; Reported quantity: 186; Lot: Lot No. K556402, Expires 2016-10. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Merit Medical Systems, Inc. is voluntarily conducting a recall due to a discrepancy between the carton and unit labeling for the Merit Laureate Hydrophilic 0.035" 260 cm Guide Wire, catalog number LWSTFS35260EX, lot number K556399 (173 units), which has a straight tip and a stiff shaft. Lot K556399 was partially mixed with another lot of Laureate 0.035" 260 cm Guide Wire, catalog number LWSTDA3
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Merit Medical Systems, Inc.
- Distributor
- Worldwide Distribution: USA Nationwide and China. No military/govt/VA consignees.
- Lot
- Lot No. K556402, Expires 2016-10
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAScout MD, REF MDS05A-01, MDS05C-01, MDS10C-01 Contents: (1) Scout MD Delivery System, Standard for use withOctober 21, 2026
- FDAEmbosphere 500-700 micrometer 2 mL Syringe Sterile Steam, REF: S420GH/EUA, S620GH/EUA; Embosphere 700-900October 7, 2026
- FDASplashwire Hydrophilic Guide Wire: REF: MSWSTDA35150 UDI-DI codes: 00884450529645 00884450529669June 24, 2026
- FDAPhilips Bridge Prep Kit REF: K12-09098C Sterile EO, Rx ONLYMay 6, 2026
- FDAMerit Laureate Hydrophilic Guide Wire, Catalog No. LWSTFS35260EX, Straight Tip Stiff Shaft, 0.035" (0.89mm),April 30, 2014
- FDAiLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations ofApril 30, 2014
Questions about this recall
Was Merit Laureate 0.035" (0.89mm) Hydrophilic Guide Wire, Angled Tip, Standard Shaft, Catalog No. LWSTDA35260EX, 260 cm (102"). recalled?
Yes. FDA published recall Z-1486-2014 on April 30, 2014. Reported hazard: Merit Medical Systems, Inc. is voluntarily conducting a recall due to a discrepancy between the carton and unit labeling for the Merit Laureate Hydrophilic 0.035" 260 cm Guide Wire, catalog number LWSTFS35260EX, lot number K556399 (173 units), which has a straight tip and a stiff shaft. Lot K556399 was partially mixed with another lot of Laureate 0.035" 260 cm Guide Wire, catalog number LWSTDA3
What should I do if I have this product?
Voluntary: Firm initiated · Letter