iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology.

Published April 30, 2014 · Recall initiated: April 3, 2014

Summary

FDA records medical device Class II recall Z-1487-2014, published April 30, 2014 and initiated April 3, 2014. Recalling firm: Boston Scientific Corporation; Status: Terminated; Reported quantity: 9 units; Lot: Model Numbers: H749ILAB120CART0 H749ILAB120INS0 H749ILAB120N2710 Serial numbers: 4219, 4900, 5701, 5827 7516, 7517, 7607 8407, 8434. the different UPNs represent minor level differences in configuration relative to mobility and no difference in any essential components or software. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Boston Scientific Corporation
Distributor
USA Nationwide Distribution in the states of NY, NC, DC, FL, PA
Lot
Model Numbers: H749ILAB120CART0 H749ILAB120INS0 H749ILAB120N2710 Serial numbers: 4219, 4900, 5701, 5827 7516, 7517, 7607 8407, 8434. the different UPNs represent minor level differences in configuration relative to mobility and no difference in any essential components or software.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology. recalled?

Yes. FDA published recall Z-1487-2014 on April 30, 2014. Reported hazard: Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.

What should I do if I have this product?

Voluntary: Firm initiated · Letter