iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology.
Published April 30, 2014 · Recall initiated: April 3, 2014
Summary
FDA records medical device Class II recall Z-1487-2014, published April 30, 2014 and initiated April 3, 2014. Recalling firm: Boston Scientific Corporation; Status: Terminated; Reported quantity: 9 units; Lot: Model Numbers: H749ILAB120CART0 H749ILAB120INS0 H749ILAB120N2710 Serial numbers: 4219, 4900, 5701, 5827 7516, 7517, 7607 8407, 8434. the different UPNs represent minor level differences in configuration relative to mobility and no difference in any essential components or software. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Boston Scientific Corporation
- Distributor
- USA Nationwide Distribution in the states of NY, NC, DC, FL, PA
- Lot
- Model Numbers: H749ILAB120CART0 H749ILAB120INS0 H749ILAB120N2710 Serial numbers: 4219, 4900, 5701, 5827 7516, 7517, 7607 8407, 8434. the different UPNs represent minor level differences in configuration relative to mobility and no difference in any essential components or software.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;October 21, 2026
- FDAImager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/038 BX5;September 30, 2026
- FDABoston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVISeptember 23, 2026
- FDALUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model NumberSeptember 9, 2026
- FDAMerit Laureate 0.035" (0.89mm) Hydrophilic Guide Wire, Angled Tip, Standard Shaft, Catalog No. LWSTDA35260EX,April 30, 2014
- FDAGE Vivid E9 Ultrasound System, a general-purpose ultrasound system, specialized for use in cardiac imaging.April 30, 2014
Questions about this recall
Was iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology. recalled?
Yes. FDA published recall Z-1487-2014 on April 30, 2014. Reported hazard: Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.
What should I do if I have this product?
Voluntary: Firm initiated · Letter